OPERATION AND INITIAL TESTING
Set the bath up as described in the earlier “Setting up Section”. For convenience it is suggested that a newly
received bath should be tested at 50°C with water as the operating liquid.
Connect the bath to the supply and switch on.
1.
Set the over temperature controller to a suitable value, see earlier.
2.
Set the set-point of the temperature controller to the desired value (50.0°C).
3.
Ensure correct operation of the bath.