MODE OF OPERATION
FAST-CAL HTM 2010
The metal block function of the Fast-Cal HTM 2010 is well suited for fast, convenient, mess free calibration of temperature
sensors.
The thermometers under test are placed into suitable holes in the block.
For the S models a calibrated reference probe should be placed into the block and the actual temperature can be read from
the temperature indicator.
For the B models an external temperature indicator should be used.
For traceable calibration the actual value of the insert temperature should be recorded along with the values from the
sensors under test.
FAST-CAL
The metal block function of the Fast-Cal is well suited for fast, convenient, mess free calibration of temperature sensors.
The Fast-Cal metal insert is placed into the calibration well.
The thermometers under test are placed into suitable holes in the metal insert.
For the S models a calibrated reference probe should be placed into the block and the actual temperature can be read from
the temperature indicator.
For the B models an external temperature indicator should be used.
For traceable calibration the actual value of the insert temperature should be recorded along with the values from the
sensors under test.
How to measure the true temperature inside the block
The controller of the Fast-Cal HTM 2010 controls and reads the temperature of the calibration volume.
How to measure the true temperature inside the block
The controller of the Fast-Cal controls and reads the temperature of the block surrounding the 25mm diameter x 148mm
deep calibration well.
Remember the following:-
The Controller
The controller is used to set a constant temperature and create an Isothermal environment for the comparison calibration
of temperature sensors.