6
ISOMET
3.12
To equalise deflection efficiency at the extremes of the scan, alternate between the minimum
and maximum desired frequencies and adjust Bragg angle to give the same efficiency for
both. (Note: the photo detector or light power meter will require repositioning for the two
angles.) .
Typical swept frequency response at 374nm
First order sweep