Iso-Tech ICM A7-4 Instruction Manual Download Page 2

Amps AC

 

Display Hold

 

19

0A

19

0A

I + (-I) = 0 

I

I

Battery Replacement

 

SPECIFICATIONS 

1-1  General Specifications 

LCD display digits :  

3 1/2 digit large scale LCD readout. 

Display count : 

2000 counts. 

Measuring rate : 

2.5 times / sec. 

Overrange display :  

“OL” is displayed for “A” and “

”    

functions, shows the real value for  
“V” function. 

Automatic power off time :  

Approximately 30 minutes after 
power on. To resume operation, 
turn the instrument off, then turn on 
again. 

Low battery indicator :  

<

 is displayed. Replace the battery 

when the 

<

 indicator appears in the  

display.

 

Power requirement :  

ALKALINE 9V battery.  

Battery life :  

ALKALINE PP3; 9V 250 hours. 

Conductor Size : 

16mm (0.63”) 

diameter. 

1-2  Environmental Conditions 

Indoor Use. 
Calibration :  

One year calibration cycle. 

Operating temperature :  

0°C ~ 30°C (

80% RH)  

30°C 

40°C (

75% RH)  

40°C 

50°C (

45%RH) 

Operating altitude : 

2000m (6562 ft) 

Overvoltage category :  

IEC 61010-1 600V CAT.

CAT.

 equipment is designed to   

protect against the transients in the  
equipment in fixed installations, 
such as distribution panels, feeders, 
short branch circuits and lighting 
systems in large buildings. 

Pollution degree : 

 

EMC : 

EN 61326-1

   

 

1-3  Electrical Specifications 

Accuracy is 

±

(% r number  

of digits) at 23°C  

±

 5°C < 80%RH. 

Temperature coefficient :  

0.2 x (Specified accuracy) 

°C,  

< 18°C, > 28°C 

 

 

Function 

Range 

Accuracy 

V

 

600 Vrms 

±(

1.5%+ 3 dgt) 

40Hz ~ 500Hz 

V

 

600 V 

±(

1% + 2 dgt) 

Overload protection : 

600 Vrms 

Input impedance : 

1M

 // less 

than 100pF

.

 

Resistance & Continuity

 

Function  

Range 

Accuracy 

 

 

2000

 

±(

1% + 2 dgt) 

Overload protection : 

600V rms

 

Max. open circuit voltage : 

3V 

Continuity check : 

The

 

internal 

sounder activates if the resistance 
of the circuit under test is less than 
25

It will then turn off if the resistance  
is increased beyond 400

.

 

AC Current

 

Function 

Range 

Accuracy 

A

 

200.0A 

±(

3% + 3 dgt)* 

 45Hz ~ 66Hz 

Overload protection : 

200A

 

* Adjacent conductor influence : 
 

<0.05 A/A

 

 

 

X

 

X

 

Shock vibration :  

Sinusoidal vibration per MIL-T- 
28800E (5 ~ 55 Hz, 3g maximum).

 

Cleaning 

Periodically wipe the case with a  
dry cloth and detergent. Do not use 
abrasives or solvents.

 

 Do not hold  

the meter beyond  
the barrier 

Tactile Barrier 

 CAT.

.600V 

with respect to  
earth for the jaw. 

Resistance

 

Continuity

 