CALIBRATION:

Internal Calibration Setting (With Division. Decimal and Capacity setting):

Press  the  ZERO  key  and  switch  on  the  scale.  The  display  will  show  "PASS  "  and
then  "000000".  The  user  should  enter  the  password  XXXX"  using  the  PCS  and
MODE keys. Now press the ZERO Key The display will show "CAL ". Press the ZERO
key  again.  The  display  will  show  "RNG  01"  use  the  following  table  to  select  the
appropriate setting of RNG

Range

RNG 01 (Single Range)

RNG 02 (Dual Range)

RNG 03 (Triple Range)

Maximum Counts

Example

30000 counts or less

1500 counts for

each range or less

1000 counts for

each range or less

300kg x 10gm

150kg x 10gm (15000 div)

300kg x 20gm (15000 div)

100kg x 10gm (10000 div)

200kg x 20gm (10000 div)

300kg x 50gm (6000 div)

LOADCELL CONNECTION TO MAIN BOARD

(E+)

(E-)

(S+)

(S-)

(SH)

Load Cell Input (Red)

Load Cell Input (Black)

Load Cell Output (Green)

Load Cell Output (White)

Shield Wire Load Cell

2. ON / OFF Switch:

 The on/off switch is provided at the bottom of the Scale

to  turn  on  and  turn  off  the  Scale  Under  most  recommended  conditions,  make
sure  to  turn  off  the  Scale  when  not  in  use  for  long  time,  and  it  is  most
recommended  that  the  power  cable  of  the  Scale  should  always  be  connected
to the mains power socket.

3.  Second  Display  Connector  :

  The  15  pin  connector  in  the  bottom  of  the  Scale

provided  for  the  second  display  connections.  The  normal  Scale  has  only  a
dummy  15  pin  connector  and  the  connections  inside  the  Scale  or  the  second
display  are  not  provided.  If  you  need  Scale  with  second  display  connection  to
the connector by default, then please order -with display model.

OPERATION

Switch On: The scale can be switched on using the on/off'switch at the bottom of
the  Scale  The  display  will  show  Start  up  message  and  Version  no.  of  the
software, and then display "0.00"

KEYS & CONTROLS

Z/T

This  key  is  used  to  zero  an  unwanted  reading  on  the  display.  This
is also used forgoing in to the calibration mode.

Hold

This key is used for holding the weight on the display,

Mode 

This  key  is  used  for  Setting  the  LOW  and  HIGH  LIMIT  in  the
scale. The key is also used as the up arrow key in the setting menu.

PCS

This key is used for counting function. This key is also used as the left
 shift key in the setting menu

"RNG  01":  Press  the  ZERO  key  the  display  will  show  "div  X",  select  the  value  of
the least count between 1,2 or 5 using the MODE key. Press the ZERO

again, the display will show "000000 ", now select the decimal point position using the
PCS  shift  key.  Press  the  ZERO  key  the  display  will  briefly  show  "CAP  "and  then
display the previously entered Capacity. Enter the new Capacity using the MODE and
PCS  Keys.  Press  the  ZERO  key  the  display  will  briefly  show  "noLoad"  and
then the internal division at no load condition. Make sure for the decimal point to stop
blinking,  and  then  press  the  ZERO  key  The  display  will  show  "LOAD  "  briefly  and
then  again  the  internal  counts.  Keep  the  calibration  weight  and  press  the  ZERO  key
the  display  will  show  "000000  ".  Now  enter  the  value  of  the  calibration  weight  kept
using  the  MODE  and  PCS  Keys.  Finally  press  the  ZERO  key  the  display  will
show the calibrated weight. Please remember that the value of the DIV, DECIMAL and
CAP should be entered remembering the maximum allowed counts are 30000 counts or
less otherwise the Scale will not accept the setting and display "ERR"

"RNG  02  ":  Follow  the  above  procedure,  only  remember  that  you  have  to  enter  the
DIV,  DECIMAL  and  CAP  twice  instead  of  once.  The  maximum  number  of  counts
allowed  for  each  capacity  range  will  be  15000  counts  or  less,  and  hence  the  setting
should be entered accordingly.

Summary of Contents for K9 Series

Page 1: ...ITA ITB Series Weighing Scales ITA ITB Series Weighing Scales Sales Service Network Scale Users Guide K9 Series ISHTAA SCALES INC INDIA www ishtaascales com Hubli Nagapattinam Alapuzha ...

Page 2: ...come to Ishtaa Scales Congratulations on your purchase of ISHTAA Electronic Weighing Scales You have chosen Right Brand to weigh your materials accurately always Every care has been taken during manufacturing to get better accuracy always UNPACKING Unpack the scale carefully and save the packing materials for future Transportation use if required Confirm the contents of the packing as per your ord...

Page 3: ... charged from time to time Ideally the battery should be charged foratleast10to12hours everytwomonths toprolongthebatterylife MAINTENANCE PleasenotethefollowingpointsofcarefortheScaletoprolongitlife l Make sure to turn off the power while installing or removing the power cable of theScalefrom themains supply l Make sure to switch of the Scale while connecting or removing the loadcell connector l M...

Page 4: ...he on off switch at the bottom of the Scale The display will show Start up message and Version no of the software andthendisplay 0 00 KEYS CONTROLS Z T This key is used to zero an unwanted reading on the display This isalso usedforgoingintothecalibrationmode Hold This keyisusedfor holdingtheweightonthedisplay Mode This key is used for Setting the LOW and HIGH LIMIT in the scale Thekeyisalso usedas...

Page 5: ...e as 0 00 the Scale will sound the beep sound below the SETLO value SET HI value is entered only If the user only enters the SET HI value and keep the SET LO value as 0 00 the Scale will sound the beep sound above the SETHI value SET LO SET HI both are entered If the user enter both SET LO and SET HI value the beep will not sound below between the SET LO and SET HI value range and thebeepwillsound...

Reviews: