Manual No. 085-3435-07
22/45
IGB/IGX Series (Overseas Specifications) Service Manual
Weighing Condition Setting Table
Address Weight
Item
Data
110A
1
Decimal point position
0: 0
1: 0.0
2
2: 0.00
3: 0.000
4
Minimum graduation
0: 1 (1-2 change in Dual range/ Multi Interval)
4: 2 (2-5-10 change in Dual range/ Multi Interval)
8
8: 5 (5-10-20 change in Dual range/ Multi Interval)
C: Invalid
110B -
Accuracy
(Resolution)
0:
1/500
1: 1/600
2: 1/750
3: 1/1000
4: 1/1200
5: 1/1500
6: 1/2000
7: 1/2500
8: 1/3000
9: 1/4000
A: 1/5000
B: 1/6000
C: 1/7500
D: 1/10000 E: 1/12000 F: 1/15000
1
2
4
Changing method
0: Single range LED stable display only (IGX)
1: 1/3000 Multi interval Stabe/Upper/Lower display (IGB)
2: Single range Upper/Lower display (IGX)
3: 1/3000 Multi interval Stabe/Upper/Lower display (IGB)
4: Single range (Body weight specification)
5: 1/3000 Dual range (A/B range change)
6: Single range (lb/kg change)
7: 1/3000 Multi interval (lb/kg change)
111A
8
1
Weight unit
(Valid only for
communications)
<111A> <111B>
0 + 0: kg (kg - lb)
8 + 0: g
0 + 1: lb
8 + 1: oz
2
4
111B
8
Filter setting
Do not change the data.
(Cut off)
(Notch) (Output rate)
0: 0.66Hz
2.50Hz
400ms
2: 0.84Hz
3.20Hz
312ms
4: 1.05Hz
4.00Hz
250ms
6: 1.31Hz
5.00Hz
200ms
8: 1.68Hz
6.40Hz
156ms
A: 2.10Hz
8.00Hz
125ms
C: 2.62Hz
10.00Hz
100ms
E: 3.35Hz
12.80Hz
78ms