S p e z i a l e l e k t r o n i k G m b H
iseg
c/o ROTECH
Email: [email protected]
Phone ++ 49 351 / 26 95 - 260
Spezialelektronik
Bautzner Landstr. 45
http://www.iseg-hv.com
Fax
++ 49 351 / 26 95 - 261
GmbH
D - 01454 Rossendorf
Germany
9
Single access CHANNEL: Actual voltage
(Read-write access)
Byte
DATA_ID
Controller (DLC = 1):
Bit
7
6
5
4
3
2
1
0
Read actual voltage at the corresponding channel
Designation
DATA
_DIR
N3 N2 N1 N0
Data
1
1
0
0
0
x
x
x
x
Description
read
Channel N
x
of 0 ... 15
⇓
Response module (DLC = 4)
Byte
DATA_ID
DATA_2
DATA_1
DATA_0
Bit
7
6
5
4
3
2
1
0
0
Designation
DATA
_DIR
N3 N2 N1 N0
LSB
Data
0
1
0
0
0
x
x
x
x
x
Description
write
Channel N
x
of 0 ... 15
Actual voltage with resolution V
O max
/ 10
∗
exp6 [V]
in DATA_2 to DATA_0
Single access CHANNEL: Actual current
(Read-write access)
Byte
DATA_ID
Controller (DLC = 1):
Bit
7
6
5
4
3
2
1
0
Read actual current at the corresponding channel
Designation
DATA
_DIR
N3 N2 N1 N0
Data
1
1
0
0
1
x
x
x
x
Description
read
Channel N
x
of 0 ... 15
⇓
Response module (DLC = 4)
Byte
DATA_ID
DATA_2
DATA_1
DATA_0
Bit
7
6
5
4
3
2
1
0
0
Designation
DATA
_DIR
N3 N2 N1 N0
LSB
Data
0
1
0
0
1
x
x
x
x
x
Description
write
Channel N
x
of 0 ... 15
Actual current with resolution I
O max
/ 10
∗
exp6 [A]
in DATA_2 to DATA_0