6700SR Refrigerated Sampler
15
Some screens display the range of acceptable numbers between pa-
rentheses. If you enter a number that is too low or high, the control-
ler beeps and erases the entry. Type a new number and continue.
Stored Data Will Be Lost
Changes to the data storage interval, module setup, hardware set-
up, or selecting a new stored program, make it necessary for the
sampler to reformat its memory. See Memory on page 97. Refor-
matting the memory clears any stored data (sample events and
readings recorded during the last running program). As a precau-
tion, the sampler displays a message (screen
If you do not need the stored data, select YES. The sampler will re-
format its memory according to your menu selection and display
the new memory capacity.
If you need to collect or view the data, select NO. The sampler will
ignore your last menu selection(s) and retain the stored data. Exit
programming by pressing the STOP key and collect the data. See
Sampling Reports on page 54.
1
6700 SAMPLER
STANDARD PROGRAMMING
For HELP at any
screen press ? key.
Turn the sampler on by pressing the
On/Off
key. Press
↵
(Enter).
This screen disappears on its own after 8 seconds.
2
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
The option PROGRAM will be blinking. Press
↵
(Enter).
3
SITE DESCRIPTION:
“FACTORY051“
CHANGE?
YES NO
The option NO will be blinking. Press
↵
(Enter).
For the purposes of this example, it isn’t necessary to change the description.
4
SELECT UNITS FOR LENGTH:
ft m
The option ft will be blinking, press
↵
(Enter).
5
DATA STORAGE
INTERVAL IN MINUTES
1
2
5
10
15
30
The “1” option will be blinking. Press
↵
(Enter).
6
NUMBER OF BOTTLES:
1 2 4 8 12 24
Select the number of bottles in your bottle kit by pressing either arrow key
until the correct number blinks. Press
↵
(Enter). For this example, select 24.
7
BOTTLE VOLUME IS
1000 ml (300-30000)
Type the volume for the bottles in your kit. For this example, 1000 is correct,
so simply press
↵
(Enter).
Entering Numbers
c
SUCTION LINE LENGTH
IS __ ft
(3-99)
•Type the length of the suction
line. Acceptable values are
shown between the
parentheses. Press enter to
accept the number.
d
INTERVAL CHANGED --
STORED DATA WILL
BE LOST! OK?
YES
NO
Example 1 Program: One Sample every 15 Minutes, One Sample in Each Bottle Using
Normal Programming Style
Summary of Contents for 6700SR
Page 2: ......
Page 4: ......
Page 18: ...6700SR Refrigerated Sampler iv...
Page 148: ...6700SR Refrigerated Sampler 130 Appendix B Material Safety Data Sheets...
Page 149: ...6700SR Refrigerated Sampler 131...
Page 150: ...6700SR Refrigerated Sampler 132...
Page 151: ...6700SR Refrigerated Sampler 133...
Page 152: ...6700SR Refrigerated Sampler 134...
Page 153: ...6700SR Refrigerated Sampler 135...
Page 154: ...6700SR Refrigerated Sampler 136...
Page 155: ...6700SR Refrigerated Sampler 137...
Page 156: ...6700SR Refrigerated Sampler 138...
Page 157: ...6700SR Refrigerated Sampler 139...
Page 158: ...6700SR Refrigerated Sampler 140...
Page 159: ...6700SR Refrigerated Sampler 141...
Page 160: ...6700SR Refrigerated Sampler 142...
Page 161: ...6700SR Refrigerated Sampler 143...
Page 162: ...6700SR Refrigerated Sampler 144...
Page 163: ...6700SR Refrigerated Sampler 145...
Page 164: ...6700SR Refrigerated Sampler 146...
Page 165: ...6700SR Refrigerated Sampler 147...
Page 166: ...6700SR Refrigerated Sampler 148...
Page 167: ...6700SR Refrigerated Sampler 149...
Page 168: ...6700SR Refrigerated Sampler 150...
Page 174: ...6700SR Refrigerated Sampler 156 Appendix D Replacement Parts REFRIGERATOR ASSEMBLY FRONT BACK...
Page 175: ...6700SR Refrigerated Sampler 157 REFRIGERATOR ASSEMBLY...
Page 176: ...6700SR Refrigerated Sampler 158 MODEL 6700 CONTROLLER INSIDE...
Page 177: ...6700SR Refrigerated Sampler 159 MODEL 6700 CONTROLLER INSIDE...
Page 178: ...6700SR Refrigerated Sampler 160 MODEL 6700 CONTROLLER OUTSIDE...
Page 179: ...6700SR Refrigerated Sampler 161 MODEL 6700 CONTROLLER OUTSIDE...
Page 180: ...6700SR Refrigerated Sampler 162 MODEL 6700 PUMP GEAR CASE ASSEMBLY...
Page 188: ...6700SR Refrigerated Sampler 170...
Page 190: ......