Doc. SIE62153
Rev. 1
Page 23/24
5.2.2 Two phase voltage generator
- Two independent voltage sources, with a common neutral.
- Type of connection: safety banana plugs.
- Output ranges, and corresponding power and resolution of AMIV-66 alone.
RANGE OUTPUTS
CONNECTION
VOLTAGE
(V)
POWER
(VA)
Z MAX
(Ohm)
RISOLUT.
1
2 X
DIRECT
0…125
80
195
1.9 mV
2
2 X
DIRECT
0…12,5
195
190
µ
V
3
2 X
DIRECT
0…1
195
19
µ
V
4
1 X
2 IN SERIES
0…250
160
390
3.8 mV
5
1X
2 IN PARALLEL
0…125
160
97
1.9 mV
OPTIONAL 300 V OUTPUT
1
2X
DIRECT
0…300
80
1125
4.6 mV
2
2X
DIRECT
0…125
80
195
1.9 mV
3
2X
DIRECT
0… 12.5
195
190
µ
V
4
1X
2 IN SERIES
0…600
160
390
9.2 mV
5
1X
2 IN PARALLEL
0…300
160
97
4.6 mV
- Output ranges, and corresponding power and resolution of AMIV-66 with DRTS3 PLUS.
RANGE OUTPUTS
CONNECTION
VOLTAGE
(V)
POWER
(VA)
Z MAX
(Ohm)
RISOLUT.
1
6 X
DIRECT
0…125
80
195
1.9 mV
2
6 X
DIRECT
0…12,5
195
190
µ
V
3
6 X
DIRECT
0…1
195
19
µ
V
4
1 X
2 IN SERIES-PAR. 0…250
320
195
3.8 mV
5
1X
4 IN PARALLEL
0…125
320
50
1.9 mV
OPTIONAL 300 V OUTPUT
1
6X
DIRECT
0…300
80
1125
4.6 mV
2
6X
DIRECT
0…125
80
195
1.9 mV
3
6X
DIRECT
0… 12.5
195
190
µ
V
4
1X
4 IN SERIES-PAR. 0…600
320
195
9.2 mV
5
1X
4 IN PARALLEL
0…300
320
50
4.6 mV
- Output frequency: from DCto 2000 Hz; transient 5 kHz.
- Waveform resolution: 28 bit (16 for the amplitude, 12 for the shape).
- Output accuracy: ± 0.1% of the output ± 0.02% of the range.
- Distortion: 0.1% total maximum, with any load.
- Automatic protection for overloads and counter-feed.
- Angle accuracy: ± 0.05°.
5.2.3 Power supply
- Mains power supply: 90 to 264 V a.c., single phase.
- Frequency: 47 to 63 Hz.