IRIS Cytofuge 12 Operator'S Manual Download Page 1

 

  

Cytofuge

®

12

  Cytocentrifuge System

 

  

 

 

 

 

Operator’s Manual 

 

 

 

 
 

 
 
 

 

 
 
 
 
 
 
 
 
 
 

A Division of IRIS International, Inc.

 

Summary of Contents for Cytofuge 12

Page 1: ...Cytofuge 12 Cytocentrifuge System Operator s Manual A Division of IRIS International Inc...

Page 2: ...2 Cytocentrifuge System Model Number M802 For in vitro diagnostic use Product Number CF12 Cytofuge 12 for 100 240 VAC 50 60 Hz StatSpin and Iris are registered trademarks of IRIS International Inc Cop...

Page 3: ...on 3 6 Overview 6 Sample Volumes 6 Pre concentration 6 Dilution 6 Section 4 7 Opening and Closing the Cover 7 Using the Removable Rotor 7 Spinning Samples 8 Programming 9 Section 5 10 Overview 10 Clea...

Page 4: ...ut during centrifugation CAUTION Do not place the cytocentrifuge on top of other equipment and do not place other objects or equipment on top of the cytocentrifuge CAUTION The cover interlock by pass...

Page 5: ...safety hood See references CAUTION The seals in the removable rotor are intended to stop gross fluid leakage only These seals should not be relied upon for protection against biohazardous contaminati...

Page 6: ...s pkg of 8 CLIP Backing Plates for Cell Concentrators pkg of 4 BP01 Filter Concentrators 2 bags of 4 each FF01 Stainless Steel Clips for Filter Concentrators 3 pkgs of 4 FFCL Install o Place the cytoc...

Page 7: ...12 completes the power on check it will give a two beep ready signal Press the STOP OPEN button to release the latch and open cover WARNING Do not use the supplied line cord outside of North America...

Page 8: ...atched and prevents access to spinning parts until the rotor has come to a complete stop Speeds from 200 rpm to 2000 rpm and cycle times from 1 minute to 99 minutes may be selected The system features...

Page 9: ...OL PANEL To change the cycle speed use the arrow keys to the right of the SPEED display Each key press will change the speed by 10 rpm Press and hold the key for 2 seconds to change the rpm in 50 rpm...

Page 10: ...face Accessories Product No Description CC03 3 well Cell Concentrators and 3 well gaskets Box of 10 CC01 1 well Cell Concentrators and 1 well gaskets Box of 10 CLIP Stainless Steel Clips for all Cell...

Page 11: ...rences Sample Volumes Use the following guidelines for sample volumes in each of the Cytofuge concentrators Max Volume Range Optimum Volume Range Filter Concentrator 50 to 300 L 100 to 300 L One well...

Page 12: ...e damage to the cytocentrifuge motor mount it is important to remove the rotor whenever the cytocentrifuge is moved or shipped to another location CAUTION The rotor CFRT 12 is intended for use with bi...

Page 13: ...l well 5 Install lid onto rotor press center button to lock 6 Place rotor into cytocentrifuge assure rotor is fully seated onto cone in center of bowl 7 Close cover and set spin cycle CAUTION Do not o...

Page 14: ...carefully remove the rotor and transport to a biological hood 3 Remove the rotor lid to access the sample concentrators Programming The Cytofuge 12 Cytocentrifuge can store up to 24 user defined progr...

Page 15: ...pt those recommended operators should check with customer service that the proposed method will not damage the equipment General Cleaning 1 Unplug cytocentrifuge from the electrical outlet 2 Open the...

Page 16: ...ne o rings or a new rotor bowl P N 01 006168 001 if necessary To replace the center seal simply pull the seal off of the rotor bowl and press on the seal as shown in the figure Checking Replacing Driv...

Page 17: ...de Table of manual to identify and correct problem Concentrators are loose and do not fit into rotor Use only Cytofuge concentrators other manufactures may look similar but only Cytofuge concentrators...

Page 18: ...fore returning for service Any instrument or accessory that has been exposed to blood or other biological materials must be cleaned prior to shipment to the manufacturer or distributor for service Thi...

Page 19: ...50 RPM sec MED 65 70 RPM sec HI 85 90 RPM sec Deceleration Fixed 1min Electrical 100 240 VAC 50 60Hz 2 5 A 300 W Physical Height 8 0 20 3 cm Width 13 0 33 0 cm Depth 16 0 40 6 cm Weight 22 lbs 10 Kg...

Page 20: ...ansmission in Health Care Settings MMWR Suppl No 2S 2S 18S 1987 3 CDC Updated US Public Health Service Guidelines for the Management of Occupational Exposures to HBV HCV and HIV and Recommendations fo...

Page 21: ...anty extended by Iris Sample Processing shall apply to any instrument that has been damaged due to misuse negligence accident or damage resulting from unauthorized repairs alterations or improper inst...

Reviews: