Changing the AC Line Fuses
The
AC
line
connection
has
two
independent
fuses.
The
fuse
holder
is
integral
to
the
power
inlet
on
the
rear
panel
of
the
SL/SLx.
To
check
or
change
fuses:
1.
Remove
the
power
cord
from
the
inlet
receptacle.
2.
Using
a
small
screwdriver,
unlatch
and
open
the
fuse
panel.
3.
Remove
and
inspect
both
fuses.
4.
If
one
or
both
are
open,
replace
them
with
new
fuses,
as
indicated
on
the
rear
panel
label.
Replacement
fuses
are
IEC
127
2A,
250V,
medium
time
delay,
5
mm
x
20
mm.
5.
If
newly
replaced
fuses
fail,
contact
IRIDEX
Technical
Support.
Resetting the Circuit Breaker
Next
to
the
power
inlet
is
a
circuit
breaker
protecting
the
power
supply
transformer
from
extended
overload.
The
circuit
breaker
senses
high
internal
operating
temperatures,
often
caused
by
high
ambient
temperatures
or
by
greatly
exceeding
typical
clinical
usage
parameters
and
overcurrent
conditions
such
as
low
line
voltage.
When
any
of
these
conditions
threaten
reliability
of
the
SL/SLx,
the
circuit
breaker
button
pops
out.
If
the
circuit
breaker
pops:
1)
Correct
any
power
input
conditions
or
allow
the
SL/SLx
to
cool.
2)
Press
the
circuit
breaker
reset
button.
3)
If
the
circuit
breaker
is
tripped
after
manually
resetting,
contact
IRIDEX
Technical
Support.
OcuLight SL/SLx Operator Manual
PN 13099D-EN
27