12

 

IRIDEX EndoProbe

® 

Handpiece, OtoProbe™ Device and FlexFiber™ Device Operator Manual

 

13103-EN Rev K

5

Safety and Compliance

To ensure safe operation and prevent hazards and unintended exposure to the laser beams, read and 

follow these instructions:

•  To prevent exposure to laser energy, except as a therapeutic application from either direct or 

diffusely reflected laser beams, always review and observe the safety precautions outlined in the 

operator manuals before using the device.

•  This device is intended for use only by a qualified physician. The applicability of the equipment 

and treatment techniques selected is your sole responsibility.

•  Do not use any device if you think it is not functioning properly.
•  Laser beams reflected from specular surfaces can harm your eyes, the patient’s eyes, or others’ 

eyes. Any mirror or metal object that reflects the laser beam can constitute a reflection hazard. 

Be sure to remove all reflection hazards near the laser. Use non-reflecting instruments whenever 

possible. Be careful not to direct the laser beam at unintended objects.

CAUTION:  Changes or modifications not expressly approved by the party responsible for 

compliance could void the user’s authority to operate the equipment.

Protection for the Physician

Eye safety filters protect the physician from backscattered treatment laser light. Integral eye safety 

filters are permanently installed in every compatible Slit Lamp Adapter (SLA) and Laser Indirect 

Ophthalmoscope (LIO). For endophotocoagulation or for Operating Microscope Adapter (OMA) use, 

a separate discrete eye safety filter assembly must be installed into each viewing path of the operating 

microscope. All eye safety filters have an optical density (OD) at the laser wavelength sufficient to 

permit long-term viewing of diffuse laser light at Class I levels. When using the handpieces, always 

wear the appropriate laser safety eyewear.
Always wear appropriate laser safety eye wear when performing or observing laser treatments with 

the unaided eye. Refer to laser console Operator Manual for laser safety eye wear minimum OD, it is 

specific per each laser console wavelength and maximum power output.

Protection for All Treatment Room Personnel

The Laser Safety Officer should determine the need for safety eyewear based on the Maximum 

Permissible Exposure (MPE), Nominal Ocular Hazard Area (NOHA), and Nominal Ocular Hazard 

Distance (NOHD) for each of the delivery devices used with the laser system, as well as the 

configuration of the treatment room. For additional information, refer to ANSI Z136.1, ANSI Z136.3, 

or European Standard IEC 60825-1.

Summary of Contents for EndoProbe Series

Page 1: ...EndoProbe Handpiece OtoProbe Device and FlexFiber Device Operator Manual ...

Page 2: ...al OcuLight EndoProbe FlexFiber and SmartKey are registered trademarks BriteLight CW Pulse DioPexy DualSense EasyFit EasyView FiberCheck FlexFiber G Probe IQ 532 IQ 577 IQ 810 LongPulse MicroPulse MilliPulse OtoProbe PowerStep SmartKey Symphony Tri Mode TruFocus and TruView are trademarks of IRIDEX Corporation All other trademarks are the property of their respective holders ...

Page 3: ...ls 5 OtoProbe Models 6 FlexFiber Models 6 Eye Safety Filters 6 Connecting the Components 7 Laser Spot Size 7 Treating Patients 7 3 Troubleshooting 9 General Problems 9 4 Maintenance 11 Cleaning Optical Components 11 5 Safety and Compliance 12 Protection for the Physician 12 Protection for All Treatment Room Personnel 12 Safety Compliance 13 Illuminating Probe Specifications 13 Illumination Testing...

Page 4: ...Contents iv 13103 EN Rev K ...

Page 5: ...seous tissue in the medical specialty of Ear Nose and Throat ENT surgery The OtoProbe is cleared for use for the particular indications of the compatible laser system to which it is attached The FlexFiber Device is intended for use in treatment of soft tissue vascular lesions of the airway and larynx The FlexFiber is cleared for use for the particular indications of the compatible laser system to ...

Page 6: ...earing the appropriate laser safety eyewear Never substitute prescription eyewear for laser safety eyewear Always handle the fiber optic cables with extreme care Do not coil the cable into a diameter less than 15 cm 6 in Turn off the laser before inspecting any delivery device components Keep the protective cap over the fiber optic connector when the delivery device is not in use Follow standard f...

Page 7: ...heal tubes from laser radiation Ignition or perforation of endotracheal tubes by the laser beam could result in serious or fatal patient complications References Reference material and additional information regarding laser safety and the prevention of endotracheal fires may be obtained from the following U S sources ANSI Z136 3 The Safe Use of Lasers in Health Care Facilities American National St...

Page 8: ...cal Support 650 962 8100 techsupport iridex com Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherlands 2797 Warranty and Service This device carries a standard factory warranty This warranty is void if service is attempted by anyone other than certified IRIDEX service personnel NOTE This Warranty and Service statement is subject to the Disclaimer of Warranties Limitation of Remedy and ...

Page 9: ...sit Ensure that the appropriate eye safety filter is installed prior to using a probe EndoProbe Models Model Description Standard straight or angled Provides intraocular laser delivery Illuminating Provides light illumination and laser delivery Stepped 45 angled Provides laser delivery probe designed for use with a trochar Adjustable and Intuitive Provides laser delivery probe designed for manual ...

Page 10: ...umber Description 15702 200 μm Laser Fiber 15703 300 μm Laser Fiber 15704 400 μm Laser Fiber 15706 600 μm Laser Fiber Eye Safety Filters Eye Safety Filter 532 nm 577 nm 810 nm Fixed Two Position Fixed Eye Safety Filter Eye protection for physician is fixed and always in place Two Position Eye Safety Filter Filter position is adjusted manually SmartKey provides feedback to laser to ensure no laser ...

Page 11: ...er exposure or injury to you your patient or others in the treatment room CAUTION Gently finger tighten the connector to the port Do not overtighten Laser Spot Size To change the spot size move the probe farther from or closer to the target Treating Patients Before treating a patient Ensure that the eye safety filter as appropriate is properly installed and that the SmartKey if used is selected En...

Page 12: ...he footswitch to deliver the treatment beam To conclude patient treatment 1 Select Standby mode 2 Record the number of exposures and any other treatment parameters 3 Turn off the laser and remove the key 4 Collect the safety eyewear 5 Remove the warning sign from the treatment room door 6 Disconnect the delivery device s 7 Disconnect the SmartKey if used 8 If the delivery device is single use disp...

Page 13: ...am Verify that the remote interlock has not been activated Verify that the aiming beam is visible Verify that the fiber switch is in the correct position for the laser system and wavelength you are using Verify that the eye safety filter is in the closed position If there is still no treatment beam contact your local IRIDEX Technical Support representative No illumination light LIO only Verify tha...

Page 14: ...ers may be too close to the tissue response threshold for consistent response Increase the laser power and or exposure duration or select a different lens Does not fit on the mounting plate OMA only Inspect and clean the mounting plate Verify that the mounting plate corresponds to your microscope Laser and viewing systems are not focussed at the same point OMA only Verify installation of a 175 mm ...

Page 15: ...or microbes and may lead to patient injury and or infection Store probe package under normal storage conditions in a dry clean well ventilated area at room temperature between 15º 25ºC 59 77ºF Cleaning Optical Components 1 Wrap a lens tissue around one end of a cotton tipped swab 2 Place several drops of 100 ethanol 100 methanol or high grade acetone on the tissue 3 Wipe the lens gently with the s...

Page 16: ...ty to operate the equipment Protection for the Physician Eye safety filters protect the physician from backscattered treatment laser light Integral eye safety filters are permanently installed in every compatible Slit Lamp Adapter SLA and Laser Indirect Ophthalmoscope LIO For endophotocoagulation or for Operating Microscope Adapter OMA use a separate discrete eye safety filter assembly must be ins...

Page 17: ...SO 15752 Section 6 2 Feature Specification Effective Aperture 0 5 mm Numerical Aperture 0 5 0 8 Light Guide Materials PMMA or Silica Illumination Testing The following information on the illumination capabilities of this device is provided in accordance with ISO 15752 Section 4 4 2 Measurement of the aphakic weighted irradiance of an illuminating EndoProbe may be made under sterile conditions Foll...

Page 18: ...entative Expiration Date Footswitch Footswitch In Footswitch Out Fuse Gauge Protective Earth Ground Illuminating Probe Decrease Increase Interval Interval with MicroPulse Laser Aperture at End of Fiber Laser Warning Illumination LOT Manufacturer Date of manufacture Off On Part Number Power Pulse Count Pulse Count Reset No ionizing Electromagnetic Radiation Read Information Remote Control Remote In...

Page 19: ...rsion Refer to Instruction Manual Booklet in blue Initial Power PowerStep Interval between Groups Number of Pulses Group Number of Steps PowerStep Power MicroPulse Power Increment Power Increment PowerStep Parameter is Locked USB Port Indicators Laser Firing Laser Preparing Speaker Screen System Brightness Latex Free Prescription Warning Replace with fuses as indicated ...

Reviews: