31
Electrostatic
Discharge
(ESD)
IEC 61000-
4-2
±8 kV
contact
±15 kV
air
±8 kV
contact
±15 kV
air
Floors should be
wood, concrete,
or ceramic
tile. If floors
are covered
with synthetic
material, the
relative humidity
should be at
least 30 %.
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-
4-8
30 A/m
30 A/m
Power frequency
magnetic
fields should
be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
Table 3:
Guidance and manufacturer’s declaration—
electromagnetic immunity
The Zio QX device is intended for use in the electromagnetic
environment specified below. The customer or the user of
the Zio QX device should assure that it is used in such an
environment.
Immunity
test
IEC 60601
test level
Com-
pliance
level
Electromagnetic
environment -
guidance