11
P2 = CPLD ID
—A2
P3 = Simmer Frequency byte
– 0x31 to
0xC7 200 KHz to 50 KHz, 5 KHz steps
P4 = Simmer Duty Cycle byte
– 0x00 to
0x14 0 to 20%, 1% steps
P5 = Pulse Stretch byte
– 0x00 to
0x64 0 to 10 us, 1 us step
P6 = Get Temp. 0x01 to 0xFF (In
Degree ‘C’) P7 = Get VSWR Status
(0x00=OK; 0x01=Fault
P8 = Get Interlock Status
(0x01=’On’; 0x00=’OFF’)
P9 = Get Temp. Warning Status
(0x01=’On’;
0x00=’OFF’) P10 = Get Laser Fault Status
(0x01=’On’; 0x00=’OFF’) P11 = Get Laser VDC -
0x0000 to 0xFFFF (85mV/count)
Packet Termination
Command Function
Cmd Data
Set Simmer Frequency
0x11 p3 = Simmer Frequency byte - 0x31 to
0xC7 200 KHz to 50 KHz, 5 KHz steps
Response Packet
0x11 Same as Get Status Response Packet
Set Simmer Duty Cycle
0x12 p1 = Simmer Duty Cycle byte - 0 to 20%, 1%
steps
Response Packet
0x12 Same as Get Status Response Packet
Set Pulse Stretch
0x13 p1 = Pulse Stretch byte
– 0 to 10 us, 1 us
step
Response Packet
0x13 Same as Get Status Response Packet
Command Function
Cmd Data
Get Laser Serial
Number
0x20