Purpose of the Device
This thermometer is intended to measure the body temperature of humans. The ear
measuring function can be used for babies who are a minimum of six months old. The
forehead function can be used for all ages. It measures the body temperature based
on heat radiating from the eardrum or from the forehead. Although this thermometer
has undergone extensive clinical tests and is considered safe and highly accurate, it is
intended for non- professional home use only.
Do not take temperature measurements on body parts other than ears and forehead, to
prevent inaccurate readings.
Do not use the ear mode when you have an ear infection.
Warranty
This device is subject to Limited Warranty. This covers any defects in materials or
workmanship under normal use during the Warranty Period. iProvèn will either
replace the product or repair the product at no charge, using new or refurbished
replacement parts. The Warranty Period of this iProvèn product is 2 years from the
date of purchase. A replacement product or product part assumes the remaining
warranty of the original product purchase. This Limited Warranty does not cover
batteries and packaging, nor any problem that is caused by conditions, malfuncti-
ons, or damage not resulting from defects in material or workmanship.
Model: Triple Mode Infrared Thermometer TMT-215
Applicable regulations and laws: ASTM 1965-98
Temperature units: ˚C/˚F, adjustable
Measurement range:
0˚C - 32.1˚C / 32˚F - 89.8˚F: ±1˚C / ±1.8˚F
32.2˚C - 42.2˚C / 89.9˚F - 108˚F: ±0.2˚C - / ±0.4˚F
42.3˚C-100˚C / 108.1˚F - 212˚F : ±1˚C / ±1.8˚F
Display resolution: 0.1˚C/0.1˚F
Latency Time: 1 second
Automatic shutdown function: ±10 + 1s
Low-voltage display function: low-voltage sign shows when the voltage is below
2.5V±0.1V
Memory function: capacity of 20 data sets
Operational conditions: ASTM 10˚C-40˚C (50˚F-104˚F)/15-95%RH
Operational atmospheric pressure range: 86-106kPa
Battery: Two 1.5V AAA type batteries
Battery life: alkaline dry battery around 20,000 measurements
Accuracy (laboratory): ±0.2°C (±0.4°F) from 36.0°C-39.0°C/96.8°F-102.2°F
±0.3°C (±0.5°F) from 32.0°C-35.9°C/89.6°F-96.7°F
±0.3°C (±0.5°F) from 39.1°C-42.2°C/102.3°F-108.0°F
iProvèn owns and reserves the rights comprised in the copyright of this document.
No part of this document may be changed, copied, reproduced, or imitated in any
form or by any means without prior written consent of iProvèn. All statements,
information, and recommendations in this document are provided “AS IS” without
warranties, guarantees or representations of any kind, either express or implied. The
information in this document is subject to change without notice. iProvèn reserves
the right of final interpretation of this document.
Error message table
Symbols
Copyright
Technical specifications
FAQs
1) How can I replace the batteries?
If the low battery symbol
is displayed, please replace the batteries.
Release the battery cover by gently sliding it down.
Remove the battery cover.
Replace both AAA- batteries according to the polarity marked on the battery chamber.
Replace the battery cover and make sure all of the pins of the cover are aligned with the
gaps. Close the battery cover by sliding it up until you hear a click.
2) How should I clean the device?
Regularly clean the thermometer probe with a damp cloth, moistened slightly with water.
You can clean the lens of the probe with a cotton swab. If you want to disinfect the sensor
probe, use a disinfecting wipe or a cotton ball dampened with a disinfectant.
You can prevent the sensor from getting stained by ensuring the ear canals are clean.
Caution: This thermometer is not waterproof! Do not immerse or expose it to water or
any other liquid other than using a slightly damp cloth.
3) How should I store the thermometer?
Store this thermometer at room temperature. The most accurate temperature measurement
is obtained when the thermometer and the user have both been exposed to the same ambi-
ent temperature for at least 30 minutes. Always store this thermometer in the enclosed soft
case to prevent it from collecting dust or being exposed to bacteria. If your thermometer
will not be used for more than 90 days, we recommend you remove the batteries and store
them separately. Do not expose the device to direct sunlight.
Safety instructions and precautions:
•
The ear measuring function of this thermometer is not suitable for children under
6 months. For children younger than 6 months we recommend using the forehead
mode or a rectal digital thermometer such as the iProvèn DT-R1221B.
•
This device is not intended to serve as a continuous monitoring device.
•
This device is not waterproof. Please keep it away from water or other liquids.
•
Do not expose the device to direct sunlight.
•
Carefully follow the instructions in the cleaning section when you need to clean and
disinfect your device.
•
Make sure children do not use the device unsupervised.
•
Please do not use the product if the temperature sensor of the thermometer shows
any sign of damage. Do not try to repair the product if it is damaged. Instead,
please contact our customer support at www.iproven.com.
•
The use of this thermometer is not intended to take the place of appropriate medi-
cal supervision.
•
This thermometer consists of high quality, precision-made parts. It is not designed
to be impact- or shock-resistant. Avoid dropping the product or subjecting it to
intense shock or vibration.
•
Remove the batteries if the thermometer will not be used for more than three
months.
Error
Meaning
Desired action
ER1 + red bac 3
beeps
Room temp below
10
˚
C / 50
˚
F or above
40
˚
C / 104
˚
F
Store device at room
temperature
ERC + red bac
3 beeps
Data processing error
Contact customer support
Low voltage signal
Battery too low
Replace batteries
Lo:
Low temperature reading
Forehead: less than 35
˚
C
(95.0
˚
F) Ear: less than
32
˚
C (89.6
˚
F) Object:
less than 0
˚
C (32.0
˚
F)
Check if you follow instruction
for accurate measurement
Hi:
High temperature reading
Forehead: more than 42.2
˚
C (108.0
˚
F) Ear: more
than 42.2
˚
C (108.0
˚
F)
Object: more than 100
˚
C
(212.0
˚
F)
Check if you follow instruction
for accurate measurement
The nitty gritty
details
SUPER
MOM!
This item is not intended to replace insight from a doctor, pharmacist, or
other licensed healthcare professional. You should not utilize this item
for self-diagnosis or for independent treatment of an illness. Immediately
contact your health-care professional if you suspect that you have a clinical
issue.
Electromagnetic interference: this thermometer contains sensitive
electronic components and should not be used in the vicinity of
electromagnetic interference (such as mobile phones or microwaves).
Please dispose of used products and batteries in accordance with local
regulations and requirements.
Symbol
Description
Type BF applied part.
Attention must be paid.
The action is prohibited.
Information about the
manufacturer.
Date of manufacture.
Consult the instructions
for use.
This product complies with
the MDD93/42/EEC requi-
rements
Waste electrical materials
should be sent to a dedi-
cated collection point for
recycling.
A personal injury or damage
to the thermometer may
occur if the thermometer is
not used correctly.
Inaccurate reading or dama-
ge to the thermometer may
occur if the thermometer is
not used correctly.
Attention