Digital488 User’s Manual
11-28-01
Introduction,
1-3
IEEE 488
CAUTION
The IEEE 488 terminal must only be used to control a non-isolated IEEE 488 system.
The common mode voltage (cable shell to earth) must be zero.
Terminal Installation Category:
Standard:
Not Applicable.
CE:
Category 1.
Implementation:
SH1, AH1, T6, TE0, L4, LE0, SR1, RL0, PP0, DC1, DT1, C0, E1.
Terminators:
Selectable CR, LF, LF-CR, and CR-LF with EOI.
Programmable:
IEEE Terminators, EOI, SRQ Mask, Port Data, Active Levels, Handshake Lines, Format and
Configuration.
Connector:
Standard IEEE 488 connector with metric studs.
General
Configuration:
Five 8-bit ports, programmable as inputs or outputs. Also included are programmable handshake lines,
data latching capability, Clear and Trigger outputs, and a Service Request (SRQ) input.
Terminal Installation Category:
Standard:
Not Applicable.
CE:
Category 1 for all terminals.
Dimensions:
188 mm deep x 140 mm wide x 68 mm high (7.39" x 5.5" x 2.68").
Weight:
1.55 kg. (3.6 lbs).
Operating Environment:
Standard:
Indoor, 0
°
to 50
°
C; 0 to 70% RH to 35°C. Linearly derate 3% RH/°C from 35 to 50°C.
CE:
Indoor use at altitudes below 2000 meters, 0
°
to 40
°
C; 80% maximum RH up to 31
°
C decreasing
linearly 4% RH/
°
C to 40
°
C.
Controls:
Power switch (external), and IEEE parameter switches (internal).
Indicators:
LED indicators for IEEE TALK, LISTEN, SRQ, ERROR, and POWER.
Power:
An external power supply is provided with the Digital488: Input is 105-125 VAC,
or 210-250 VAC; 50/60 Hz, 10 VA maximum. The external power supply 9 VDC output
is to be connected to the Digital488 power input marked: 10 VDC MAX @ 500 mA.
WARNING
Do not use this interface outdoors. The interface is intended for indoor use only.
Outdoor conditions could result in equipment failure, bodily injury, or death.
CAUTION
Do not connect AC power line directly to the Digital488. Direct AC connection will
damage equipment.
Summary of Contents for Digital488
Page 21: ...2 10 Getting Started 11 19 01 Digital488 User s Manual...
Page 37: ...3 16 Command Descriptions 7 17 01 Digital488 User s Manual...
Page 48: ...Digital488 32 OEM Component Layout Digital488 User s Manual 916895 Service Information 5 5...
Page 49: ...Digital488 OEM Component Layout 5 6 Service Information 916895 Digital488 User s Manual...
Page 51: ...5 8 Service Information 916895 Digital488 User s Manual...
Page 61: ...B 8 Notes...
Page 62: ...Appendix C Digital488 OEM Mechanical Dimensions Board Dimensions C 1...
Page 63: ...IEEE Connector Mounting and Hole Punch Out C 2...