![IOtech Digital488/80A User Manual Download Page 59](http://html1.mh-extra.com/html/iotech/digital488-80a/digital488-80a_user-manual_2088030059.webp)
CAUTION
Because the Configuration (C) command is used, the following programming
example(s) will not work properly if the HVCX1 option is installed with a conflicting
hardware configuration. To be safe, remove the HVCX1 board and replace it with the
jumper board to run the example(s).
EXAMPLE 5: Command Conflict Error.
PRINT#1,"OUTPUT08;G0C5P1X"
Line 1:
Configure all 5 ports as outputs; select Port 1 only.
PRINT#1,"OUTPUT08;D123456ZX"
Line 2:
Send too much data for one port.
PRINT#1,"OUTPUT08;E?"
Line 3:
Query the error status (ERROR LED should be on).
PRINT#1,"ENTER08",
INPUT#2,A$,
PRINT A$
Line 6:
Display shows:
E3-Conflict Error
(ERROR LED should
be off).
52 Digital488/80A Commands
967695
Digital488/80A User's Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com