Carefully unscrew the black plastic Filter Cover
using both hands placed on the knurled sections for
better grip.
3.
Check the colour of the indicators.
4.
Check the sealing O-rings, if damaged replace
immediately.
5.
If the result is ok then screw the Filter Cover back
on.
To replace the Moisture Filter follow the instructions below:
1)
Place the GasClam 2 on a table with the head down.
2)
Carefully unscrew the black plastic Filter Cover.
3)
Using both hands, put the thumbs beneath the filter and push
it up and out of the connecting barb and nozzle.
4)
Leaning the new filter to right and first connect the tube
(Tygon) to the barb.
5)
Push the tube firmly down on the barb.
6)
Lean the filter back upright making sure the other inlet sits
down on the nozzle.
7)
Push the filter completely down firmly from the top.
8)
Run the “Leak Test”
from the software to check that the
system is correctly sealed
9)
Check the sealing O-rings, if damaged replace immediately.
10)
If the result is ok then screw the Filter Cover back on.
User Serviceable Parts
Battery compartment O
–ring:
When changing the batteries always inspect the O-ring on the Battery
Cover Plate, if there is any sign of damage replace.
Moisture Filter
The GasClam 2 must never be operated without a Moisture Filter! Follow the
advice regarding Moisture Filter management and the instructions on how to
change the filter.
Filter Compartment O
–rings:
When inspecting the Moisture Filter check O-ring 1 and 2, if there are any signs
of damage replace the corresponding O-ring immediately.
Snorkel
The snorkel should be checked regularly, if there is any damage replace immediately.
Rubber Collar
Inspect the collar regularly, if there is any signs of damage replace immediately.