high over time. To fix this issue, the membrane cartridge must be replaced. It is crucial that the probe
not remain in water as this can damage the sensor and electronics within. Use the following steps to
change the membrane cartridge:
1)
Carefully unscrew the protective guard from the probe body. Slide the membrane cartridge
away from the probe body, being careful not to damage or stress the membrane tubing which
connects to the stainless-steel sensor port in the center of the probe body.
2)
Pinch the head of the membrane tube; carefully separate the membrane tubing from the sensor
port and set the now completely detached membrane assembly aside.
3)
Install the new membrane cartridge by pressing the tubing over the sensor port until the tubing
is within 1 mm from the base of the sensor port. Slide the membrane cartridge over the
temperature sensor and sensor port.
4)
Inspect the tubing where it meets the sensor port to ensure the inner tubing has not collapsed.
5)
Screw the protective cover to the probe body, over the membrane cartridge.