2-3
(1)
On the
3150(M) MRI Monitor
, turn the Power Switch to the ON position.
(2)
Once the 3150(M) MRI Monitor has come on line, turn the front panel Power
Switch on the Remote Monitor to the ON position.
(3)
After the Remote Monitor is on line, verify that the signal is received from
the 3150(M) MRI Monitor
(4)
Verify that the lower right hand box on the Display Screen is labeled
REMOTE, both squares inside the box are together and the STATUS box is
Green.
(5)
Depress (press and release) the
NIBP START/STOP
Control Key on the
Remote Monitor.
(6)
Verify that an NIBP Determination is initiated on the 3150(M) MRI Monitor.
(7)
Verify that the NIBP Determination is taken and displayed on the 3150(M)
MRI Monitor and Remote Monitor.
(8)
Verify P2 Waveform and Values are the same on both the 3150(M) MRI
Monitor and Remote Monitor.
(9)
With P2 still ON, verify that the SpO2 Value is shown with no waveform and
is the same on both the 3150(M) MRI Monitor and Remote Monitor.
(10)
Turn P2 OFF and verify SpO2 waveform is now displayed.
(11)
Verify EtCO2 Waveform and Values are the same on both the 3150(M) MRI
Monitor and Remote Monitor.
b.
Remote Transmission Failures.
If the Remote Transmission System fails to operate
properly, consult the following guidelines for possible causes.
(1)
Transmission System will not Link.
An inability to link is most often cause
by a loose antenna or a faulty power up. If the system will not link first check
the Antenna Connections to ensure a proper tight mounting and, if the system
still fails to link, cycle the power by turning the power OFF, waiting 5
seconds and then turning the power ON. If Remote Transmission is still not
possible, contact a Qualified Service Representative for further assistance.
(2)
Transmission System Drops Out.
A Drop Out is any momentary loss of
communication between the monitors that comprise this system. Drop Outs
can be caused by any number of variables present in the local environment.
If Drop Outs occur, consult the following guidelines:
(a)
Ensure the Antenna Connections are tight and secure.
(b)
Attempt to position the monitor's so that the base of the 3150(M)
MRI Monitor has an unobstructed line of sight to the remote
monitor's antenna. If reception improves then an object within the
previous line of sight was interfering with transmission.
(c)
If possible, move the 3150(M) MRI Monitor and/or remote monitor
to different locations. If reception improves it is possible that the
Summary of Contents for Magnitude 3150M
Page 1: ......
Page 2: ......
Page 4: ......
Page 22: ......
Page 32: ......
Page 38: ......
Page 52: ......
Page 54: ......
Page 64: ......
Page 66: ......
Page 67: ...D 1 APPENDIX D DECLARATION OF CONFORMITY...
Page 68: ...D 2...
Page 73: ...F 1 APPENDIX F LIST OF SYMBOLS...
Page 74: ......
Page 75: ...NOTES...
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......