![Invacare XPO100 Service Manual Download Page 8](http://html.mh-extra.com/html/invacare/xpo100/xpo100_service-manual_2082635008.webp)
SECTION 1—IMPORTANT SAFEGUARDS
XPO
2
™
8
Part No 1150748
Radio Frequency Interference
This equipment has been tested and found to comply with EMC limits specified by IEC/
EN 60601-1-2. These limits are designed to provide a reasonable protection against
electromagnetic interference in a typical medical installation.
Other devices may experience interference from even the low levels of electromagnetic
emissions permitted by the above standards. To determine if the emissions from the XPO
2
are causing the interference, turn the XPO
2
Off. If the interference with the other device(s)
stops, then the XPO
2
is causing the interference. In such rare cases, interference may be
reduced or corrected by one of the following measures:
• Reposition, relocate, or increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different from that to which the
other device(s) is connected.