Quick Guide
Identifying and
r
epairing
f
aults
There
are
audio
and
visual
alarms
present
on
the
control
unit.
Problem / Alarm
Cause
Solution
Mattress CPR hose disconnected.
Connect CPR hose connector, lock it in place.
Power cable and fuse has been checked, control
unit does not operate.
Send control unit back to Invacare for repair.
Mattress not inflating
(not alternating
properly).
Alternating system
/ Timing failure
indication
Major leak in air cell or complete air insert.
Replace leaking air cell.
CPR hose or tube connectors kinked or split.
Unkink or replace split CPR hose or tube
connectors.
Not alternating, rotor failure.
Send control unit back to Invacare for repair.
No air (control unit failure).
Send control unit back to Invacare for repair.
Control unit off.
Check power source, turn unit on.
No power /
Power failure
indication
Power cord disconnected.
Connect power cord and ensure the power
source is on.
No power in the power outlet.
Switch on the power outlet (if pole switch
available).
Have the power outlet repaired by an electrician.
Power outage.
Wait until the power source has power.
Fuse blown.
Change fuse on power inlet connector with spare
fuse or identical replacement only (consult a
trained engineer if you are unsure how to change
a fuse).
Disconnection of CPR (connection hose).
Connect hose properly.
Disconnection of connector tubes to air cells in
air insert.
Check individual air cells in insert are correctly
connected to connector tubes.
Pressure failure / Low
pressure indication
Kinked connection hose
Make sure that there are no kinks or bends in
the hose.