Control unit menu display
K
J
G
H
I
C D E F
A
B
Quick Guide
Invacare®
Soft
form Active
®
2 with RX pump
Pos.
Description
Function
A
Mute button
The
relevant
audible
/
visible
indicator
turns
on
when
low
pressure,
power
failure
or
alternate
failure
is
recognised.
To
mute
the
audible
signal,
press
the
Mute
button.
The
visible
indicator
will
flash
u ntil
the
problem
is
solved.
If
the
problem
indicated
has
not
been
solved
within
10
minutes,
the
indicator
will
sound
again.
Press
the
Mute
button
to
mute
the
audible
indication,
the
indicator
will
not
sound
again.
Refer to
Identifying and repairing faults
on page 3 of this guide
.
B
System in cycle / Setup indicator
(Operative LED)
Green LED indicator flashes whilst system is setting up – Solid green LED
indicates normal operational.
C
Alternating system / Timing failure
indicator
Red LED indicator flashes and audible buzzer sounds if the system
recognises a cycle time problem.
D
Power failure indicator
Red LED indicator flashes and audible buzzer sounds if the system has
no power connected. This could be in the event of a power failure, for
example caused by power cord unplugged or power off when control
unit is working. When power is restored, control unit will automatically
start working.
E
Pressure failure / Low pressure
indicator
Red LED indicator flashes and audible buzzer sounds if the system
recognises a low pressure failure.
Operation
1.
The
power
indicator
will
flash
and
control
unit
will
enter
static
mode
initially
until
pressure
reaches
15
mmHg.
Then
control
unit
will
enter
alternate
mode
and
power
indicator
will
remain
on.
The
cycle
time
is
set
at
10
min,
and
the
pressure
is
set
at
60
±
3
mmHg.
The first inflation of new mattress: the indicator function (visible and audible indicators) will activate if the mattress
does not inflate completely within 15 minutes.
Operation failures
•
During
control
unit
operation,
low
pressure
indicator
will
activate
within
1
minute
if
air
hoses
are
disconnected
or
air
cells
are
broken.
•
In
the
case
of
abnormal
alternate
or
no
alternate,
the
audible
signal
will
be
activated
and
alternate
failure
indicator
will
flash.