Control  unit  menu  display

K

J

G

H

I

C D E F

A

B

Quick Guide

Invacare®

 

Soft

form Active

®

 2 with RX pump

Pos.

Description

Function

A

Mute button

The

 

relevant

 

audible

 

/

 

visible

 

indicator

 

turns

 

on

 

when

 

low

 

pressure,

 

power

 

failure

 

or

 

alternate

 

failure

 

is

 

recognised.

 

To

 

mute

 

the

 

audible

 

signal,

 

press

 

the

 

Mute

 

button.

 

The

 

visible

 

indicator

 

will

 

flash

 

u ntil

 

the

 

problem

 

is

 

solved.

 

If

 

the

 

problem

 

indicated

 

has

 

not

 

been

 

solved

 

within

 

10

 

minutes,

 

the

 

indicator

 

will

 

sound

 

again.

 

Press

 

the

 

Mute

 

button

 

to

 

mute

 

the

 

audible

 

indication,

 

the

 

indicator

 

will

 

not

 

sound

 

again.

 Refer to 

Identifying and repairing faults

 on page 3 of this guide

.

B

System in cycle / Setup indicator
(Operative LED)

Green LED indicator flashes whilst system is setting up – Solid green LED
indicates normal operational.

C

Alternating system / Timing failure
indicator

Red LED indicator flashes and audible buzzer sounds if the system
recognises a cycle time problem.

D

Power failure indicator

Red LED indicator flashes and audible buzzer sounds if the system has
no power connected. This could be in the event of a power failure, for
example caused by power cord unplugged or power off when control
unit is working. When power is restored, control unit will automatically
start working.

E

Pressure failure / Low pressure
indicator

Red LED indicator flashes and audible buzzer sounds if the system
recognises a low pressure failure.

Operation

1.

The

 

power

 

indicator

 

will

 

flash

 

and

 

control

 

unit

 

will

 

enter

 

static

 

mode

 

initially

 

until

 

pressure

 

reaches

 

15

 

mmHg.

 

Then

 

control

unit

 

will

 

enter

 

alternate

 

mode

 

and

 

power

 

indicator

 

will

 

remain

 

on.

 

The

 

cycle

 

time

 

is

 

set

 

at

 

10

 

min,

 

and

 

the

 

pressure

 

is

 

set

 

at

 

60

 

±

 

3

 

mmHg.

The first inflation of new mattress: the indicator function (visible and audible indicators) will activate if the mattress
does not inflate completely within 15 minutes.

Operation failures

During

 

control

 

unit

 

operation,

 

low

 

pressure

 

indicator

 

will

 

activate

 

within

 

1

 

minute

 

if

 

air

 

hoses

 

are

 

disconnected

 

or

 

air

 

cells

 

are

 

broken.

In

 

the

 

case

 

of

 

abnormal

 

alternate

 

or

 

no

 

alternate,

 

the

 

audible

 

signal

 

will

 

be

 

activated

 

and

 

alternate

 

failure

 

indicator

 

will

 

flash.