SoftAIR Alternating Mattress System

The SoftAIR Excellence is a alternating mattress system 
that provides pressure application and release to patients 
who are vulnerable to or suffer from pressure ulcers. It is 
designed to replace an existing mattress and can be used 
on both standard and profiling bed frames.

Digital Control Unit 

T

he Control Unit provides the air supply to the mattress. 

• 

It is controlled via a touch panel with integrated 
digital display. The alarm sounds when pressure 
fails or power is interrupted. The Alarm Mute 
button silences the alarm for a maximum of 20 
minutes.  The alarm resumes if cause of failure is 
not resolved. The alarm will sound for up to two 
hours following an interruption to power. As soon as 
power is restored, the battery memory back up will 
restore the latest pressure/ user settings.  

• 

The Control Unit is fitted with an ‘on board’ battery. 
The battery powers the alarm sounder in the event 
of a total power failure or power interruption.

• 

The buttons on the control panel adjust the eight 
comfort level settings.

• 

The system will automatically revert to Alternation 
Mode after 20 minutes in Static Mode. See page 8 
for further instructions.

• 

The Alarm LED indictor and Alarm Mute completes 
the profile.

The visible and audible alarm function has a number of 
indications depending on the cause of the failure.

On the side of the control unit are four male air connectors 
for connecting the handle.

The rapid release handle includes a cap which can be 
inserted into the handle during transportation. When in 
transportation mode, tests have proven that the system 
will stay inflated for up to 30 hours with no significant loss of 
pressure. Whilst in transportation mode, the mattress will 
remain in static mode (i.e. cells will not alternate). For this 
reason, Invacare recommend that the alternating mode is 
resumed once transportation is complete.

The mains supply to the Control Unit can be easily 
disconnected and is designed to detach if tugged too 
hard, protecting the internal wiring.    

SoftAIR Excellence 

The SoftAIR Excellence Mattress comprises of 19 high 
density cells which all feature a permanently inflated 
internal cell to prevent the patient “bottoming out” in 
the event of low pressure due to incorrect settings, 
electrical or cell failure. 

This system includes three static head cells to provide 
static “pillow” support for optimum user comfort, while 
air pressure in the other 16 cells is alternated over a 
10-12 minute cycle. This provides regular periods of 
pressure reduction to aid blood and lymphatic flow to 
vulnerable tissue. 

The SoftAIR Excellence also includes an independent 
heel zone of five low pressure micro cells for 
individualised therapy to this sensitive area.

In addition, two permanently inflated side formers assist 
in lateral support for the user and carer while increasing 
the effectiveness of the one in three alternation cycle. 
The hinged mattress base conforms to movements of 
a profiling bed .

Summary of Contents for SoftAir EXCELLENCE

Page 1: ...SoftAirEXCELLENCE250mmAlternatingMattressSystem User Manual...

Page 2: ...mportant to read this manual and understand the operating instructions and safety precautions Failure to do so could result in patient injury and or damage to the product If you have any questions ple...

Page 3: ...tem Installation 7 Control UnitActivation 7 OPERATION 8 Control Unit Panel 8 Mattress Function 9 Alarm Functions 10 Removal Transport Function 11 TROUBLESHOOTING 12 CLEANING 14 MAINTENANCE 16 TECHNICA...

Page 4: ...se the power cord supplied with the system Exposure of the electronic Control Unit to any liquid while it is plugged in could result in a severe electrical hazard Always use fuses of the same rating a...

Page 5: ...trolunitarefourmaleairconnectors forconnectingthehandle Therapidreleasehandleincludesacapwhichcanbe insertedintothehandleduringtransportation Whenin transportationmode testshaveproventhatthesystem wil...

Page 6: ...e that may have occurred during transportation Any damaged or missing parts must be reported to your local Invacare Ltd sales office immediately Please refer to the back page for contact details SoftA...

Page 7: ...ide rails as straps will tear a Position the Control Unit by placing the hooks over the foot board of the bed or side rails b Connect the handle to the Control Unit Ensure the air hose does not kink b...

Page 8: ...back toAlternation Mode after 20 minutes E Pressure Arrow Buttons Press the arrows to increase or decrease the pressure setting Eight pressure settings are available from soft to hard 18mmHg to 60mmHg...

Page 9: ...wered 5 Repeat the Bottoming Out test after pressure has been lowered In the event of a system malfunction the alarm will activate and pressure LEDs will flash Establishing Pressure inclined position...

Page 10: ...s and corresponding Pressure Setting LED displays are illustrated below Operation Display Alarm Signal Initial Failure Mattress has failed to reach minimum operational pressure within 50 minutes Low P...

Page 11: ...nce transportation is complete Regularly perform a Bottoming Out test to ensure the patient is appropriately supported as air pressure is released from all internal static cells as well as alternating...

Page 12: ...oses Check the CPR tag is attached and both sealing connectors are firmly secure 3 Check all air hoses along the inside of the mattress each should be firmly connected Check each air cell is securely...

Page 13: ...t the air hoses to reduce pressure reconnect when pressure has decreased 3 Check for twists in the air hoses between the mattress and Control Unit 4 Switch on Power Patient is sinking or bottoming out...

Page 14: ...n heat dry cycle If there are no visible signs of body fluids and or substances on the top cover the top cover should be sanitised 1 Apply an intermediate level disinfectant or a solution of Sodium Hy...

Page 15: ...to prevent scratches 4 Carefully remove the air filter cover Remove and discard the filter and fit with a new filter there may be a small locking screw use a small Phillips Head screwdriver to remove...

Page 16: ...2000mm Width 830mm 880mm Height 250mm including side formers Weight 9 5kg 10 8kg Control Unit Dimensions Length 290mm Width 170mm Height 270mm Weight 3 2kg Cell Material 0 15mm TPU film laminated on 2...

Page 17: ...prior to return Returns should include information describing the problem and or requested repair and be sent to your local Invacare Ltd sales office by prepaid transportation Invacare Ltd will return...

Page 18: ...ark Tel 45 58 85 27 22 Fax 45 58 85 43 86 Invacare Norway Invacare AS Grensesvingen 9 Postboks 6230 Etterstad N 0603 Oslo Norway Tel 47 22 57 95 00 Fax 47 22 57 95 01 Invacare Sweden Invacare AB Fager...

Reviews: