11.1.2 Installing the Reliant Scale
WARNING!
-The patient and sling MUST be removed from the lift during ALL installation procedures.
1. Position the load cell assembly of the Reliant Scale into the hanger bar bracket. Refer to Detail A.
Use a 1/4-inch nylon washer.
2. Secure the Reliant Scale to the hanger bar bracket with a new locknut and the hardware removed in
. Refer to Detail A for washer orientation.
Ensure the shaft of the shoulder bolt passes through both sides of the hanger bar bracket.
3. Tighten the locknut onto the shoulder bolt until the locknut bottoms out.
4. Insert the hanger bar pin with two nylon washers through the hanger bar. Refer to Detail B.
5. Align the mounting holes in the hanger bar pin with the mounting holes in the load cell assembly. Refer to Detail B.
6. Secure the hanger bar pin to the load cell assembly with the provided mounting screw and locknut. Securely tighten. Refer to
Detail B.
WARNING!
-After ANY adjustments, repair or service and BEFORE use, make sure all attaching hardware is securely tightened;
otherwise, injury or damage may occur.
1078987-O
62
11 Accessories
Summary of Contents for Reliant 450
Page 16: ...3 Product Labeling Fig 3 1 1078987 O 16 3 Product Labeling...
Page 37: ...Fig 7 3 1078987 O 37 7 Patient Lifting...
Page 61: ...Fig 11 1 61 1078987 O Invacare Reliant 450 600...
Page 71: ...Notes 1078987 O 71...
Page 72: ...Notes 72 1078987 O Invacare Reliant 450 600...
Page 86: ...3 tiquettes sur le produit Fig 3 1 1078987 O 86 3 tiquettes sur le produit...
Page 106: ...Fig 7 3 106 1078987 O Invacare Reliant 450 600...
Page 129: ...Fig 11 1 129 1078987 O Invacare Reliant 450 600...
Page 138: ...Remarques 1078987 O 138...
Page 139: ...Remarques 139 1078987 O Invacare Reliant 450 600...
Page 140: ...Remarques 1078987 O 140...
Page 154: ...3 Etiquetado de Producto Fig 3 1 1078987 O 154 3 Etiquetado de Producto...
Page 174: ...Fig 7 3 1078987 O 174 7 Levantando al Paciente...
Page 211: ...Notas 1078987 O 211...