10 MAINTENANCE
45
Manual/Electric Portable Patient Lift
Replacing the Electric Actuator
1.
Remove the nut, washer and shoulder bolt that secure the electric actuator to the mast mounting bracket.
2.
Rest the boom on your shoulder and remove the nut, bolt, plastic bushing and pinch guard from the boom mounting bracket.
3.
Remove the electric actuator assembly.
4.
Reverse steps for installation.
FIGURE
10.2
Maintaining the Manual/Hydraulic Pump - Replacing a Manual/Hydraulic Pump -Replacing the Electric Actuator
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
Boom
Pinch Guard
Bushing
Mast
Boom
Pinch
Guard
Bushing
Mast
Manual/Hydraulic