Installing Front Shroud
1.
Insert tabs at bottom of front shroud (1) into slots in
bottom of frame (3).
2.
Secure bottom of front shroud (1) to frame using two
screws (2). Do not overtighten.
6.10.3 Replacing Rear Shroud
•
Slotted screwdriver
•
Phillips screwdriver
Removing Rear Shroud
1.
Remove two screws (2) that secure rear shroud (4) to
frame (3).
2.
Remove rear shroud (4).
Fig. 6-17
Installing Rear Shroud
1.
Insert tabs at bottom of rear shroud (4) into slots in
bottom of frame (3).
2.
Secure rear shroud (2) to frame (3) using two screws
(2). Do not overtighten.
6.11 Controls
6.11.1 Replacing Power Module
When replacing the power module or remote, take into account the final selection of the drive program, refer to
Drive Program Selection After Component Replacement, page 39
Different combinations of power modules and remotes can be fitted to the mobility device. Power modules are described
in earlier revisions of this service manual or in the LiNX service manual.
CAUTION!
Any changes to drive program can affect driving characteristics and tipping stability of mobility device.
– Changes to drive program may only be carried out by qualified technicians.
– Invacare can only give a warranty for safe mobility device driving behaviour – especially tipping stability –
for unaltered standard drive programs.
All power modules are delivered with a standard drive program. If you have made any customer-specific
modifications to the drive program, these must be adapted after the installation of the new power module.
•
8 mm socket wrench
•
To adapt the drive program: programming software and LiNX service manual, available from Invacare.
1568396-C
37