Invacare® Platinum® Mobile
General
Alarm
Name
Alarm
Code
Alarm Code Name
Display Message
Possible Reason
Solution
Low Oxygen
Purity
(Running)
0xa20a
LPA_lowOxygenPurity
The display shows:
"Low Oxygen See
Manual"
1.
The concentrator
is producing less
than 85% O2 and
greater than 73%
O2 beginning 15
minutes after the
power-on state.
a. Unit running
without cannula
connected -or-
b. Inlet airflow
blockage -or-
c. Outlet filter
perforated -or-
d. Sieve bed
has been
compromised
-or-
e. Oxygen sensor
is not properly
calibrated.
1.
Ensure cannula is
connected while
unit is operating.
2.
Clean air intake
filter screen.
3.
Change outlet
filter.
4.
If alarm persists,
contact Invacare
for service.
28
1187375-B