Pressure Relief Mechanism
Operational at:
241 kPa ± 34.5 kPa (35 psi ± 5 psi)
Change in maximum recommended
flow when back pressure of 7 kPa
is applied:
0.7 L/min
Filters:
Cabinet (2), Outlet HEPA and Compressor Inlet
Safety System:
Current overload or line surge shutdown.
High temperature compressor shutdown.
Internal High Temperature shutdown.
High Pressure Alarm with compressor shutdown.
Low Pressure Alarm with compressor shutdown.
Battery Free Power Loss Alarm.
SensO
2
Oxygen System.
Possible Obstruction Alert.
Width:
46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)
Height:
67.0 cm ± 1 cm (26 3/8 in ± 3/8 in)
Depth:
36.5 cm ± 1 cm (14 3/8 in ± 3/8 in)
Weight:
24.0 kg ± 1 kg (53 lb ± 2 lb)
Shipping Weight:
27.2 kg ± 1 kg (60 lb ± 2 lb)
Operating Ambient Temperature
and Humidity:
10°C to 35°C (50°F to 95°F) at up to 60% relative humidity
Cabinet:
Impact Resistant flame-retardant plastic cabinet that conforms to UL 94 V-0
No AP/APG:
Not suitable for use in the presence of a flammable anaesthetic mixture.