Invacare Platinum IRC5LX Operator'S Manual Download Page 9

9

Flow Range:

Average Power

Consumption:

Pressure Relief Mechanism

Operational at:
Change in maximum

recommended flow

when back pressure

of 7kPa is applied:

Filters:

Safety System:

Width:

Height:

Depth:

Weight:

Shipping Weight:

SPECIFICATIONS

SPECIFICATIONS (CONTINUED)

1/2 - 5 L/min.
For flowrates less than 1 L/min.,

we recommend the use of the

Invacare Pediatric Flowmeter IRCPF16.

400W

35 psi + 3.5 psi (241 kPa + 24.1 kPa)

0.7 L/min.

Cabinet (2), Outlet HEPA and

Compressor Inlet.

Current overload or line surge shutdown.

High temperature compressor shutdown.

High Pressure Alarm w/compressor shutdown.

Low Pressure Alarm w/compressor shutdown.

Battery Free

 

Power Loss Alarm.

Sens

O

2

® Oxygen System 

(

Sens

O

2

® Model)

18 3/8” +/- 3/8” (46.7 cm ± 1 cm)

26 3/8” +/- 3/8” (67.0 cm ± 1 cm)

14 3/8” +/- 3/8” (36.5 cm ± 1 cm)

51 lbs. +/- 1 lb (23.2 kg +/- .5 kg)

56 lbs. +/- 1 lb (25.4 kg +/- .5 kg)

Summary of Contents for Platinum IRC5LX

Page 1: ...S MANUAL MUST BE GIVEN TO THE END USER USER BEFORE USING THIS CONCENTRATOR READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE DISTRIBUIDOR ENTREGUE ESTE MANUAL AL USUARIO FINAL USUARIO ANTES DE OPERAR EL CONCENTRADOR LEA TOTALMENTE ESTE MANUAL Y CONSÉRVELO COMO REFERENCIA PARA EL FUTURO Invacare 5 with SensO2 Model IRC5LXO2 Invacare 5 con SensO2 Modelo IRC5LXO2 Invacare 5 Model IRC5LX Invacare 5 Model...

Page 2: ...CAUTIONS AND INSTRUCTIONS CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY RESULT SPECIAL NOTES WARNING CAUTION notices used in this manual apply to hazards or unsafe practices which could result in personal injury or property damage NOTICE THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE ...

Page 3: ...ibed 13 Oxygen Tubing 14 Power Switch 15 Flowrate 15 SensO2 Oxygen Purity Indicator If your Unit Features the O2 Sensor 15 Initial Startup of the Concentrator 16 Explanation of Indicator Lights 16 Oxygen Purity Indicator If your Unit Does Not Features the O2 Sensor 17 MAINTENANCE 17 Cleaning the Cabinet Filters 17 Cabinet 18 Humidifier 18 TROUBLESHOOTING GUIDE 19 TABLE OF CONTENTS TABLE OF CONTENT...

Page 4: ...ntly displayed A spontaneous and violent ignition may occur if oil grease or greasy substances come in contact with oxygen under pressure These substances MUST be kept away from the oxygen concen trator tubing and connections and all other oxygen equipment DO NOT use any lubricants unless recommended by Invacare For optimum performance Invacare recommends that each concentrator be on and running f...

Page 5: ...n concentrator RADIO FREQUENCY INTERFERENCE Most electronic equipment is influenced by Radio Frequency Interference RFI CAUTION should be exercised with regard to the use of portable communications equipment in the area around such equipment TO REDUCE THE RISK OF BURNS ELECTROCUTION FIRE OR INJURY TO PERSONS WARNING ALWAYS unplug this product IMMEDIATELY after using Avoid using while bathing If co...

Page 6: ...en concentrators or oxygen therapy devices Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance In certain circumstances oxygen therapy can be hazardous Invacare recommends that you seek medical advice before using this machine This equipment is to be used as an oxygen supplement and is not considered life supportin...

Page 7: ...16 Venture Homefill II home oxygen compressor IOH200 Power Cord Outlet Fitting NOTE This outlet fitting is to be used only for filling oxygen cylinders with the Venture HomeFill II home oxygen compressor The outlet fitting does not affect concentrator performance Refer to the HomeFill II owner s manual part number 1100873 for connection and operating instruc tions When not in use the plug provided...

Page 8: ...tion levels From 6 000 ft 1828 meters to 13 129 ft 4000 meters below 90 efficiency ALL 5LXO2 5LX MODELS 95 6 to 87 at 1 2 to 5 L min Stated concentration levels achieved after initial warm up period approximately 30 minutes 5 0 5 psi 34 5 3 45 kPa SPECIFICATIONS SensO2 Alarm Thresholds O2 Purity Below 73 73 to 85 Over 85 Indicators RED Indicator Light Continuous Audible Alarm Sieve GARD Compressor...

Page 9: ...n we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 400W 35 psi 3 5 psi 241 kPa 24 1 kPa 0 7 L min Cabinet 2 Outlet HEPA and Compressor Inlet Current overload or line surge shutdown High temperature compressor shutdown High Pressure Alarm w compressor shutdown Low Pressure Alarm w compressor shutdown Battery FreePower Loss Alarm SensO2 Oxygen System SensO2 Model 18 3 8 3 8 46 7 cm 1...

Page 10: ...han Ambient 4 F 2 C Impact Resistant flame retardant plastic cabinet that conforms to UL 94 V0 DOUBLE INSULATED PRODUCT ETL certified complying with UL 1097 U S models CSA certified to C22 2 No 68 Canadian models No extension cords No closer than 3 inches from any wall furniture draperies or similar surfaces Crush Proof Tubing 7 ft 2 1m cannula with a maximum 50 ft 15 2 m extension DO NOT PINCH 20...

Page 11: ... damage is evident notify the Carrier or your local dealer 2 Remove all loose packing from the carton 3 Carefully remove all the components from the carton FIGURE 1 PACKAGING HANDLING NOTE Unless the Oxygen Concentrator is to be used immediately retain containers and packing materials for storage until use of the concentrator is required INSPECTION 1 Examine exterior of the Oxygen Concentrator for...

Page 12: ...venient Your concentrator can be easily rolled from room to room on its casters Your Oxygen Concentrator will perform best when operated under the following conditions Usage in environments other than those described may result in the need for increased equipment maintenance The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and or fumes RECOMMENDED...

Page 13: ...at your concentrator is at least 3 three inches away from walls draperies or furniture to assure sufficient air flow Avoid deep pile carpets and heaters radiators or hot air registers PLUG IN POWER CORD 1 Plug in power cord to an electrical outlet CONNECT HUMIDIFIER IF SO PRESCRIBED 1 Fill humidifier with distilled water to the level indicated by the manufacturer OPERATING INSTRUCTIONS Humidifier ...

Page 14: ...tructions which accompany the accessory 4 Attachoxygentubingfromthehumidifierbottletooxygenoutletconnectoron the oxygen concentrator Oxygen Tubing Oxygen Outlet Connector WARNING For optimum performance Invacare recommends that concentrator be operated a minimum of 30 minutes during any use Shorter periods of operation may adversely affect the life of the product Filter Access Door FIGURE 4 ATTACH...

Page 15: ... the L min line prescribed FIGURE 6 FIGURE 6 ADJUSTING THE FLOWRATE PLATINUM 5 WITH SENSO2 SHOWN Ball Flow knob POWER SWITCH 1 Press power switch to the ON position The power light I O located on the control panel will illuminate a GREEN light The alarm will beep three 3 times indicating that the system is functioning properly FIGURE 5 FLOWRATE 1 Turn the flowrate knob to the setting prescribed by...

Page 16: ...URE 7 GREEN LIGHT Normal Operation YELLOW LIGHT Call Supplier IMMEDIATELY You may continue to use the concentrator unless instructed otherwise by your supplier Be certain that BACKUP OXYGEN is nearby RED LIGHT Total unit shutdown Switch IMMEDIATELY to BACKUP OXYGEN supply Call Supplier IMMEDIATELY GREEN LIGHT with YELLOW LIGHT FLASHING Call Supplier IMMEDIATELY Oxygen sensor malfunctioning you may...

Page 17: ...pending on environmental conditions FIGURE 10 NOTE Environmental conditions that may require more frequent cleaning of the filters include but are not limited to high dust smoking air pollutants etc OXYGEN PURITY INDICATOR IF YOUR UNIT DOES NOT FEATURE THE O2 SENSOR LX MODELS INDICATOR LIGHT EXPLANATION FIGURE 8 RED LIGHT Total Unit Shut Down Switch IMMEDIATELY to a BACK UP OXYGEN supply Call Supp...

Page 18: ...midifier every day 2 Wash it in soapy water and rinse it with a solution of ten 10 parts water and one 1 part vinegar 3 Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier WARNING DO NOT overfill Humidifier Bottle MAINTENANCE 2 Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly FIGURE 10 3 Dry the filter...

Page 19: ... Do not use extension cords Move to another electrical outlet or circuit 3 Call Provider for Service YELLOW or RED LIGHT Illuminated Only applies to SensO2 models 1 Low oxygen purity 2 Kinked or blocked tubing cannula or humidifier 3 Flowmeter set at 1 2 L min or less 4 Unit overheating due to blocked air intake 5 Insufficient power at outlet 6 Internal repairs required TROUBLESHOOTING GUIDE SYMPT...

Page 20: ...rescribed L min Flow Rate EVERY 90 DAYS Check Oxygen Concentration UNITS WITHOUT SensO2 ANNUALLY OR DURING PREVENTIVE MAINTENANCE SCHEDULE Clean Replace Cabinet Filter s Check Outlet HEPA Filter Check Compressor Inlet Filter Check Oxygen Concentration SensO2 MODELS Check Power Loss Alarm AS REQUIRED Clean Heat Exchanger Replace Exhaust Muffler Rebuild Top End of Compressor Refer to Preventive Main...