5.
Attach the humidifier bottle adapter
E
to the humidifier
bottle
B
by turning the wing nut
G
on the humidifier bottle
counterclockwise until it is securely fastened.
6.
When placing the humidifier bottle in the compartment
F
on
the concentrator, attach the humidifier bottle/adapter tube
H
to
the oxygen outlet connector
I
on the concentrator.
When using tubing connections longer than 7 ft (2.1 m)
in length, position the humidifier as close to the patient
as possible to allow for maximum humidification output.
7.
Attach the patient’s nasal cannula supply tube to the humidifier
bottle outlet
J
.
8.
After assembly, ensure that oxygen is flowing through the nasal
cannula.
9.
Replace the filter access door on the side of concentrator by
snapping it back into the plate groove.
6.3.2
Turning the Concentrator On
1.
Press power switch
A
to On position.
All the panel lights and the audible alarm will come
on for one second, confirming that the indicators are
functioning properly.