background image

Invacare® Platinum™10L
Oxygen Concentrator

IRC10LX

en

HomeFill® System Compatible

User Manual

This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.

Summary of Contents for PLATINUM 10L

Page 1: ...Platinum 10L Oxygen Concentrator IRC10LX en HomeFill System Compatible User Manual This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future r...

Page 2: ...on from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted Making Life s Experiences Possible is a tra...

Page 3: ...2 Select a Location 17 6 3 Setting Up the Concentrator 18 6 3 1 Setting Up the Humidifier 18 6 3 2 Turning the Concentrator On 20 6 3 3 Flowrate 21 6 3 4 Units without SensO2 21 7 Maintenance 22 7 1 S...

Page 4: ...avoided may result in property damage or minor injury or both Gives useful tips recommendations and information for efficient trouble free use Refer to the following table for the meaning of symbols m...

Page 5: ...ualified technician 1 2 Intended Use The intended function and use of the Invacare Platinum Oxygen Concentrator is to provide supplemental oxygen to patients with respiratory disorders by separating n...

Page 6: ...For Use The intended function and use of the Invacare Platinum Oxygen Concentrator is to provide supplemental oxygen to patients with respiratory disorders by separating nitrogen from room air by way...

Page 7: ...Safety 2 1 Label Locations A Serial Number Label is located on the resonator intake assembly B C Specification and Double Insulation Label is located on the back of the concentrator at the base 111835...

Page 8: ...Invacare Platinum 10L Oxygen Concentrator IRC10LX D 8 1118353 M 00...

Page 9: ...age From Fire Textiles oil or petroleum substances grease greasy substances and other combustibles are easily ignited and burn with great intensity in oxygen enriched air and when in contact with oxyg...

Page 10: ...ed with this device You should contact your local home care supplier for recommendations on which of these devices will be best for you They should also give you advice on the proper usage maintenance...

Page 11: ...Make sure concentrator is off when not in use WARNING Risk of Injury or Damage Invacare oxygen concentrators are specifically designed to minimize routine preventive maintenance To prevent injury or...

Page 12: ...a reasonable protection against electromagnetic interference in a typical medical installation Other devices may experience interference from even the low levels of electromagnetic emissions permitted...

Page 13: ...compressor IOH200 The HomeFill outlet fitting I is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor The outlet fitting does not affect concentrator performance Re...

Page 14: ...Invacare Platinum 10L Oxygen Concentrator 3 2 Pneumatic Diagram 14 1118353 M 00...

Page 15: ...nula mask tubing humidifier etc used to deliver oxygen to the patient need to include a means to reduce the propagation of fire in the accessories for the safety of the patient and others If a commerc...

Page 16: ...he carton Unless the oxygen concentrator is to be used IMMEDIATELY leave concentrator in its packaging for storage until use is required 5 2 Inspection 1 Inspect examine exterior of the concentrator f...

Page 17: ...imal performance Locate and position the concentrator in a well ventilated space so that the air intake and the air exhausts are not obstructed NEVER block the air openings of the concentrator or plac...

Page 18: ...t if input and output connections are reversed When using tubing connections longer than 7 ft 2 1 m in length position the humidifier as close to the patient as possible to allow for maximum humidific...

Page 19: ...ure prior to use CAUTION Risk of Damage Confirm the humidifier cap is not cross threaded on the humidifier bottle 3 Insert a flathead screwdriver in the plate groove C on the top edge of the filter ac...

Page 20: ...s longer than 7 ft 2 1 m in length position the humidifier as close to the patient as possible to allow for maximum humidification output 7 Attach the patient s nasal cannula supply tube to the humidi...

Page 21: ...be triggered This is a rapid beeping of the audible alarm Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle After rated flow is restored to 1 L min the Pot...

Page 22: ...et DANGER Risk of Injury or Damage Liquid will damage the internal components of the concentrator To avoid damage or injury from electrical shock Turn Off the concentrator and unplug the power cord be...

Page 23: ...tor without the filter installed or with a dirty filter There are two cabinet filters located on each side of the cabinet 1 Remove the filter A and clean as needed Environmental conditions that may re...

Page 24: ...r Only qualified personnel should perform cleaning and disinfection of the oxygen concentrator and accessories between patients Follow these instructions to eliminate possible pathogen exchange betwee...

Page 25: ...EN PATIENTS UNITS WITHOUT SensO2 Every 6 months of continuous use Equivalent to 4 380 hours Check Oxygen Concentration green indicator light Clean Replace Cabinet Filter s Refer to Cleaning the Cabine...

Page 26: ...individual product marking for disposal 8 2 Wear and Tear Invacare reserves the right to ask for any item back that has an alleged defect in workmanship See Warranty that shipped with the product for...

Page 27: ...nit overheating due to blocked air intake 2 Insufficient power at outlet 3 Internal repairs required 1 Do one or both of the following a Remove and clean cabinet filters b Move oxygen concentrator at...

Page 28: ...evels 87 to 95 6 at 2 L to 10 L min Concentration levels achieved after initial warm up period approximately 30 minutes Maximum Outlet Pressure 9 psi 0 5 psi 62 0 kPa 3 45 kPa Flow Range 2 L to 10 L m...

Page 29: ...0 lbs 2 lbs 27 2 kg 1 kg Operating Ambient Temperature and Humidity 50 F to 95 F 10 C to 35 C at up to 60 relative humidity Cabinet Impact Resistant flame retardant plastic cabinet that conforms to UL...

Page 30: ...t 15 m of Crush Proof Tubing DO NOT pinch Time of Operation Up to 24 hours per day Recommended Storage and Shipping Temperature 20 F to 150 F 29 C to 65 C at up to 95 relative humidity Measurement unc...

Page 31: ...ormal wear and tear or shipping charges Invacare Corporation s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement Routine maintenance items su...

Page 32: ...IVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE...

Page 33: ...Notes...

Page 34: ...Notes...

Page 35: ...Notes...

Page 36: ...Tel 440 329 6000 Tel 800 832 4707 Technical Services Tel 440 329 6593 Tel 800 832 4707 www invacare com Manufacturer Invacare Corporation 2101 E Lake Mary Blvd Sanford FL 32773 Tel 407 321 5630 111835...

Reviews: