10 Technical data
10.1
Typical Product Parameters
Electrical Requirements:
230 VAC + 10, –15% (253 VAC/195.5 VAC), 50 Hz
Rated Current Input:
1.5 A
Sound Level:
40 dBA max
Altitude:
Up to 6,000 ft (1828 m) above sea level without degradation of
concentration levels. Not recommended for use above 6,000 ft (1828
m).
Atmospheric pressure range: 101.33 kPa — 81.22 kPa
Oxygen Output Concentration Levels:
87% to 95.6% at 0.5 to 5 L/min
93% ± 3% at 2 L/min
Concentration levels achieved after initial warm-up period
(approximately 30 minutes)
Maximum Outlet Pressure:
34.5 kPa ± 3.45 kPa (5 psi ± 0.5 psi)
Low Flow Alarm:
0.5 to 5 L/min (maximum). For flowrates less than 1 L/min, we
recommend the use of the Invacare Pediatric Flowmeter Accessory
(IRCPF16AW).
Potential Obstruction Alert:
0 L/min to 0.5 L/min
The concentrator detects a condition that may indicate a potential
obstruction of the output oxygen. Rapid audible beeping alert (this alert
is deactivated when accessories are connected). May be associated
with flow setting of 0.5 L/min or less.
Power Consumption:
Perfecto
2
AW - 300 W at 5 L/min, 280 W at 3 L/min
Perfecto
2
VAW - 320 W at 5 L/min