Invacare NordBed Essential Service Manual Download Page 12

Invacare® NordBed™ Essential

IMPORTANT!

Wrong fluids or methods can harm or damage
the product.

– All cleaning agents and disinfectants used

must be effective, compatible with one
another and must protect the materials they
are used to clean.

– Never use corrosive fluids (alkalines, acid etc.)

or abrasive cleaning agents. We recommend
an ordinary household cleaning agent such as
dishwashing liquid, if not specified otherwise
in the cleaning instructions.

– Never use a solvent (cellulose thinner,

acetone etc.) that changes the structure of
the plastic or dissolves the attached labels.

– Always make sure that the product is

completely dried before taking into use again.

For cleaning and disinfection in clinical or
long-term care environments, follow your in-house
procedures.

5.3.2 Cleaning Intervals

IMPORTANT!

Regular cleaning and disinfection enhances
smooth operation, increases the service life and
prevents contamination.
Clean and disinfect the product

– regularly while it is in use,
– before and after any service procedure,
– when it has been in contact with any body

fluids,

– before using it for a new user.

5.3.3 Cleaning Instructions

IMPORTANT!

– The product does not tolerate cleaning

in automatic washing systems, with
high-pressure cleaning equipment or steam.

Electrical components

Method: Wipe off with a wet cloth or soft brush.

Max. temperature: 40 °C

Solvent/chemicals: Ordinary household cleaning
agents and water.

Drying: Wipe dry with a soft cloth

Metal components

Method: Wipe off with a wet cloth or soft brush.

Max. temperature: 40 °C

Solvent/chemicals: Ordinary household cleaning
agents and water.

Drying: Wipe dry with a soft cloth

Metal components may also be cleaned with
pressurized water, but not high pressure or steam.

Wood

(including textile straps on side rails, if existing)

Method: Wipe off with a wet cloth or soft brush.

Max. temperature: 40 °C

Solvent/chemicals: Ordinary household cleaning
agents and water.

Drying: Wipe dry with a soft cloth

Textiles

(including upholstery and mattresses)

See attached label on each product.

5.3.4 Disinfection instructions

Information on recommended disinfectants
and methods can be found on https://vah-
online.de/en/for-users.

In domestic care

Method: Follow the application notes for the used
disinfectant and wipe-disinfect all accessible surfaces.

Disinfectant: Ordinary household disinfectant.

Drying: Allow the product to air-dry.

In clinical and long term care

Follow your in-house disinfection procedures and only use
the disinfectants and methods specified therein.

5.4 Control Unit

The control unit is placed underneath the mattress support
platform near the foot end of the bed.

1.

To open the control unit, turn the lock 90 degrees
counter-clockwise and remove the lid.

2.

To close the control unit, place the lid and turn the
lock 90 degrees clockwise.

The sockets inside the control unit are labeled with
symbols, indicating where to connect the respective plugs.
Additionally, the sockets for the actuators are labeled with
colors matching the color coding on the actuator cables.

12

1657008-A

Summary of Contents for NordBed Essential

Page 1: ...Invacare NordBed Essential en Bed Service Manual PROVIDER Keep this manual The procedures in this manual MUST be performed by a qualified technician...

Page 2: ...ation duplication or modification in whole or in part is prohibited without prior written permission from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare...

Page 3: ...1 1 Installing side rails 8 4 1 2 Installing the Freya Embla side rail 8 4 1 3 Installing the Ane side rail 8 4 1 4 Installing the Torill side rail 9 4 2 Backup battery 9 5 Servicing 11 5 1 General se...

Page 4: ...ee the user manual For more information about the product for example product safety notices and product recalls contact your Invacare representative See addresses at the end of this document 1 2 Symb...

Page 5: ...egional differences refer to your local Invacare catalog or website for available accessories or contact your local Invacare representative See addresses at the end of this document IMPORTANT Some rep...

Page 6: ...es 2 Align the slots in the bed end with the tabs on the Sanne bracket and place the bed end onto the tabs 3 Push down firmly and ensure that the bed end is fully seated in the tabs 4 Rotate the locki...

Page 7: ...cm between the bed end and the mattress bow to avoid entrapment of fingers Always make sure to tighten the screws for the mattress support extension The bed can be extended by 10 cm in the foot end of...

Page 8: ...into the bed end guideways Make sure that all side rail bars are above the locking pin and properly locked Make sure that the straps between the bars are not damaged or loose Mounting 1 Lift the uppe...

Page 9: ...ide rails of the same length can be mounted on each side of the bed The installation of the long and the short version of the Torill side rail is similar 1 Raise the side rail to its high position and...

Page 10: ...the batteries to be tested at least once a year Batteries are not damaged by continuous connection to the mains Exchanging the batteries WARNING Environmental Hazard Disposal of batteries This product...

Page 11: ...ttress support section actuators running properly q Cables correctly wired and undamaged q Electric plugs undamaged q Cable insulation and actuator housings no cracks or damages q Damaged coating repa...

Page 12: ...Wipe dry with a soft cloth Metal components Method Wipe off with a wet cloth or soft brush Max temperature 40 C Solvent chemicals Ordinary household cleaning agents and water Drying Wipe dry with a so...

Page 13: ...Backup battery optional C not in use covered with a blind plug D Height adjustment actuator lower half E Mains power cable F Height adjustment actuator upper half G Tight section actuator H Back sect...

Page 14: ...nd routed correctly Make sure cables do not touch the floor and cannot block the castors Make sure cables cannot get squeezed by moving parts during operation For the correct routing of the cables bet...

Page 15: ...agnetic key A in the slot on the side of the hand control 2 Press and hold button B and C for 3 seconds to initiate the system reset and hold the buttons pressed until the following sequence is comple...

Page 16: ...353 1 810 7085 ireland invacare com www invacare ie United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ Tel 44 0 1656 776 222 Fax 44 0 1656 776 220 uk invacare com www in...