HOMEFILL II COMPRESSOR
COMPRESSOR
PROBLEM
SOLUTION
No indicator lights on when
1. Turn compressor OFF.
compressor turned on.
2. Check that power cord is properly plugged into compressor and
wall outlet
.
3. Turn the compressor ON.
4. Clean air filters. Refer to PROCEDURE 6 - MAINTENANCE.
5. If performing STEPS 1-4 does not cause any indicator lights to come
on, contact your Home Health Care Provider or Invacare for service.
Red Light ON
1. Check the cylinder gauge to see if it is FULL.
CYLINDER FULL -
Remove the cylinder and turn the unit off.
Perform the CYLINDER PREFILL INSPECTION with a cylinder that
is NOT FULL.
CYLINDER NOT FULL -
Proceed to STEP 2.
2. Check that the cylinder is connected properly. Turn the compressor
OFF. Remove the cylinder and reinstall. Turn the compressor ON.
3. If cylinder is not full and is connected properly, the compressor may have
internal failure. Contact your Home Health Care Provider or Invacare.
Compressor cycles ON and OFF 1. Check the concentrator liter flow setting is at 3 L/min. or LESS.
OR
If necessary, adjust the liter flow setting (refer to the concentrator
O2 BELOW NORMAL
owners manual). Wait at least 20 minutes for the GREEN light to
(YELLOW) and FILLING (GREEN)
appear. If it does not appear, proceed to STEP 2.
lights are alternating
2. The concentrator may need service, contact your Home
OR
Health Care Provider or Invacare.
O2 BELOW NORMAL (YELLOW)
light goes ON and OFF
O2 BELOW NORMAL (YELLOW) 1. Ensure the concentrator has warmed up for at least 20 minutes.
light stays ON
If it has not been 20 minutes, turn the compressor OFF until the
concentrator warms up. If the YELLOW light still remains on
proceed to STEP 2.
2. Inspect the connection between the compressor and the
concentrator for damage or pinching. If damaged, replace. If pinched,
straighten tubing. Turn the compressor OFF for at least 10 seconds.
Turn the compressor ON. If the YELLOW light still remains on
proceed to STEP 3.
3. The concentrator may need service, contact your Home Health Care
Provider or Invacare.
Fill times exceed maximum
1. Ensure the cylinder regulator is set to 0 OFF.
recommended fill times shown
2. Check the recommended fill times shown in CYLINDER FILL TIMES.
in CYLINDER FILL TIMES.
3. Check that the cylinder is connected properly. Turn the compressor
OR
OFF. Remove the cylinder and reinstall. Turn the compressor ON.
GREEN filling light stays on.
4. Check the compressor/concentrator line for proper connection
and/or damage (ie. - leaks or kinks).
5. Check the concentrator liter flow setting is at 3 L/min. or LESS. If
necessary, adjust the liter flow setting (refer to the concentrator
owners manual). If the GREEN filling light does not go off, the
compressor and/or concentrator may need service. Contact your
Home Health Care Provider or Invacare.
Excessive noise when compressor 1. Call Home Health Care Provider or Invacare for service.
is ON
TROUBLESHOOTING
PROCEDURE 5
This Procedure includes the following:
Compressor
Cylinder
TROUBLESHOOTING