background image

15

VENTURE

®

 HOMEFILL™ II COMPRESSOR

COMPRESSOR

PROBLEM

SOLUTION

No indicator lights on when

1. Turn compressor OFF.

compressor turned on.

2. Check that power cord is properly plugged into compressor and

wall outlet

.

3. Turn the compressor ON.

4. Clean air filters. Refer to PROCEDURE 6 - MAINTENANCE.

5. If performing STEPS 1-4 does not cause any indicator lights to come

on, contact your Home Health Care Provider or Invacare for service.

Red Light ON

1. Check the cylinder gauge to see if it is FULL.

CYLINDER FULL -

 Remove the cylinder and turn the unit off.

Perform the CYLINDER PREFILL INSPECTION with a cylinder that

is NOT FULL.

CYLINDER NOT FULL - 

Proceed to STEP 2.

2. Check that the cylinder is connected properly. Turn the compressor

OFF. Remove the cylinder and reinstall. Turn the compressor ON.

3. If cylinder is not full and is connected properly, the compressor may have

internal failure. Contact your Home Health Care Provider or Invacare.

Compressor cycles ON and OFF 1. Check the concentrator liter flow setting is at 3 L/min. or LESS.

OR

If necessary, adjust the liter flow setting (refer to the concentrator

O2 BELOW NORMAL

owner’s manual).  Wait at least 20 minutes for the GREEN light to

(YELLOW) and FILLING (GREEN)

appear. If it does not appear, proceed to STEP 2.

lights are alternating

2. The concentrator may need service, contact your Home

OR

Health Care Provider or Invacare.

O2 BELOW NORMAL (YELLOW)

light goes ON and OFF
O2 BELOW NORMAL (YELLOW) 1. Ensure the concentrator has warmed up for at least 20 minutes.

light stays ON

If it has not been 20 minutes, turn the compressor OFF until the

concentrator warms up. If the YELLOW light still remains on

proceed to STEP 2.

2. Inspect the connection between the compressor and the

concentrator for damage or pinching. If damaged, replace. If pinched,

straighten tubing. Turn the compressor OFF for at least 10 seconds.

Turn the compressor ON. If the YELLOW light still remains on

proceed to STEP 3.

3. The concentrator may need service, contact your Home Health Care

Provider or Invacare.

Fill times exceed maximum

1. Ensure the cylinder regulator is set to “0” OFF.

recommended fill times shown

2. Check the recommended fill times shown in CYLINDER FILL TIMES.

in CYLINDER FILL TIMES.

3. Check that the cylinder is connected properly. Turn the compressor

OR

OFF. Remove the cylinder and reinstall. Turn the compressor ON.

GREEN filling light stays on.

4. Check the compressor/concentrator line for proper connection

and/or damage (ie. - leaks or kinks).

5. Check the concentrator liter flow setting is at 3 L/min. or LESS. If

necessary, adjust the liter flow setting (refer to the concentrator

owner’s manual). If the GREEN filling light  does not go off, the

compressor and/or concentrator may need service. Contact your

Home Health Care Provider or Invacare.

Excessive noise when compressor 1. Call Home Health Care Provider or Invacare for service.

is ON

TROUBLESHOOTING

PROCEDURE 5

This Procedure includes the following:
Compressor
Cylinder

TROUBLESHOOTING

All manuals and user guides at all-guides.com

Summary of Contents for IOH200

Page 1: ...0 DEALER HOMECARE PROVIDER THIS MANUAL MUST BE GIVEN TO THE USER OF THE COMPRESSOR REFER TO THE INFORMATION ON PAGE 3 OF THIS MANUAL USER BEFORE USING THE COMPRESSOR READ THIS MANUAL AND SAVE FOR FUTU...

Page 2: ...RY OR DAMAGE MAY OCCUR SPECIAL NOTES Check all parts for shipping damage In case of damage DO NOT use Contact Invacare for further instruction WARNING CAUTION notices used in this manual apply to haza...

Page 3: ...OUBLESHOOTING 15 COMPRESSOR 15 CYLINDER 16 PROCEDURE 6 MAINTENANCE 17 CLEANING THE COMPRESSOR FILTERS 17 CABINET 17 CYLINDER FILL TIMES 18 WARRANTY 19 DEALER INFORMATION NOTE Invacare recommends leavi...

Page 4: ...L LOCATIONS FRONT LEFT SIDE Keep all connections free of oil and grease to avoid violent ignition See Owner s Manual before cleaning WARNING 1075220 PART NO DO NOT REMOVE UNIT COVER This unit contains...

Page 5: ...ge per sists place the cylinder outdoors and notify your Home Health Care Provider or Service Representative of this condition DO NOT allow oil from your hands or other source to come into contact wit...

Page 6: ...at stable surface such as a table TRANSPORTING THE COMPRESSOR FIGURE 1 WARNING NEVER transport a compressor with a cylinder connected to or stored on the compressor otherwise injury or damage can occu...

Page 7: ...nd secure to the back of the concentrator with the fastening straps FIGURE 3 CONNECTING THE COMPRESSOR TO THE CONCENTRATOR FIGURE 2 INSTALLING THE POWER CORD Electric Connector Power Cord Compressor R...

Page 8: ...r repeat this checklist CYLINDER PREFILL INSPECTION FIGURE 1 WARNING ONLY use cylinders that have the Invacare Venture HomeFill II connection and the label shown in FIGURE 1 All cylinders must be insp...

Page 9: ...ONTENTS INCLUDING EMER GENCY PROCEDURES CAUTION IMPROPER FILLING OR ABUSE OF THIS CYLINDER OR FAILURE TO HEED THIS CAUTION MAY CAUSE SERIOUS INJURY OR LOSS OF LIFE DO NOT ALTER OR MODIFY CYLINDER OR R...

Page 10: ...ent WARNING DO NOT drop oxygen cylinders Use two 2 hands when handling transporting oxygen cylinders Otherwise injury or damage may occur 4 Position the cylinder in the compressor cradle 5 Align the c...

Page 11: ...The WAIT YELLOW light will be lit while the compressor warms up B AFTER 3 MINUTES The FILLING GREEN light will be on while the cylinder is filling The FULL GREEN light will be on when the cylinder is...

Page 12: ...l 1 Press the compressor power switch to the OFF position 2 Grasp the outer ring of the fillport connector and PUSH DOWN 3 Lift up on the cylinder to remove from the fillport connector 4 Perform one 1...

Page 13: ...ging the L min setting on the flow knob will affect the dose of oxygen delivered DO NOT readjust the L min setting unless directed by your physi cian or therapist 3 To turn the flow of oxygen OFF turn...

Page 14: ...ON Cylinder filling stops Cylinder is finished filling full Cannot fill another remove cylinder cylinder until compressor FULL light is on in 10 mins is turned OFF and ON or again Cylinder is not atta...

Page 15: ...and OFF O2 BELOW NORMAL YELLOW 1 Ensure the concentrator has warmed up for at least 20 minutes light stays ON If it has not been 20 minutes turn the compressor OFF until the concentrator warms up If...

Page 16: ...nds continue call Home Health Care Provider or Invacare Regulator is ON but O2 is not 1 Refill cylinder being delivered 2 Check that cannula is straight and not pinched If damaged replace 3 Call Home...

Page 17: ...udes the following Cleaning the Compressor Filter Cabinet MAINTENANCE CLEANING THE COMPRESSOR FILTER FIGURE 1 WARNING DO NOT operate the compressor without the filter installed NOTE All compressors ha...

Page 18: ...M6 AND ML6 SIZECYLINDERS CONCENTRATOR FLOWRATETO PATIENT UPTO3L MIN 1HR 25MIN 1HR 50MIN CYLINDER FILLING TIMES NORMAL MAXIMUM C M9 SIZE CYLINDER CONCENTRATOR FLOWRATETO PATIENT UPTO3L MIN 2HR 20MIN 2...

Page 19: ...back page during normal working hours Upon receiving notice of an alleged defect in a product Invacare will issue a serialized return authorization It shall be the responsibility of the purchaser at t...

Page 20: ...enture are registered Elyria Ohio USA Mississauga Ontario trademarks of Invacare Corporation 44036 2125 L4Z 4G4 Canada HomeFill and Yes you can are 800 333 6900 800 668 5324 trademarks of Invacare Cor...

Reviews: