SECTION 2—CONCENTRATOR VERIFICATION AND MAINTENANCE
17
Connecting the Compressor to the Concentrator
NOTE: For this procedure, refer to FIGURE 2.1 on page 17.
1. Insert one end of the interconnect hose into the outlet fitting on the back of the
concentrator (Detail “A”).
2. Insert the opposite end of the interconnect hose into the inlet fitting on the side of the
compressor (Detail “B”).
NOTE: The metal tabs on the concentrator outlet fitting and the compressor inlet fitting will pop
out with an audible “click” when the end of the interconnect hose is properly installed (Detail “C”).
3. For models that DO NOT have coiled tubing, loop any excess interconnect hose and
secure to the back of the concentrator with the fastening straps.
Concentrator Verification and Maintenance
Insert End of
Interconnect Hose into
Inlet/Outlet Fitting
Inlet/Outlet Fitting
Metal
Tab
Interconnect
Hose (or
Coiled tubing)
Compressor
Inlet Fitting
Power
Cord
Outlet Fitting
CONCENTRATOR
TOP VIEW
DETAIL “A”
DETAIL “C”
DETAIL “B”