Invacare® AVIVA™ RX series
Component
Comments
C
Fender
D
Nut
E
Washer
F
Fender mount
G
Wheel lock mount
H
Motor-gearbox unit
I
Width adapter
Optional component.
J
Bolts
Available in two different lengths. If a width adapter is retrofitted, the bolts must be
replaced.
Removing Unit
1.
Turn off electronics.
2.
Unplug motor cable at power module.
3.
Remove wheel including wheel hub.
4.
Remove fender. See
6.7.6 Replacing Fender, page 29
.
5.
Loosen and remove nuts
D
and washers
E
.
6.
If fitted, remove mounts for wheel lock
G
and fender
F
.
7.
Pull off unit
H
of bolts
J
.
Fig. 6-26
8.
If fitted, remove adapter
I
.
Installing Unit
1.
Check that sealing ring and unit are correctly mounted. Nut must be located on the outside and rotation of motors
must follow the specifications.
2.
Install parts in reverse order.
3.
Plug in motor cable to power module.
4.
Perform calibration process. See in LiNX service manual.
5.
Test all functions.
6.9.2 Replacing or rotating motor/gearbox unit sealing ring
CAUTION!
Risk of crushing
The motor / gearbox unit is very heavy. Risk of injury to hands.
– Pay attention to heavy weight.
38
1654647-B