![Invacare 5310IVC Owner'S Operator And Maintenance Manual Download Page 35](http://html1.mh-extra.com/html/invacare/5310ivc/5310ivc_owners-operator-and-maintenance-manual_2082254035.webp)
SECTION 5—TROUBLESHOOTING
Part No 1114836
35
IVC
™
Bed Series
Electrical Inspection and Maintenance
❑
Inspect all electrical bed components for damage or excessive wear (i.e. cracked or
broken housings, or worn components).
❑
Check pendant, power and motor cords for chafing, cuts or excessive wear.
❑
Make sure all plugs are fully attached and free of damage.
❑
Make sure cable lock on junction box is properly positioned and locked
❑
Check all functions:
A. Ensure head raises and lowers properly.
B. Ensure foot raises and lowers properly.
C. Ensure bed ends raise and lower properly.
Pass/Fail
❑
PASS: Bed and all components have passed all of the above inspection criteria and the
bed is ready for placement into a client’s home.
❑
FAIL: Bed, or any component, has failed any of the above inspection criteria. Tag the
bed or component with a complete description of the failure(s) and have the bed
serviced.