InterTest iShot UV/White LED Light Source Kit Manual Download Page 7

7

www.intertest.com

Operating Manual

Operating Manual

www.ushio.com

7

Maintenance and Cleaning

Repair

1.  Turn OFF light source by rotating intensity control knob counterclockwise 

until the switch clicks off. 

2.  Completely disconnect power supply from both light source and mains 

(power plug into ac outlet).

3.  Wait for ~5 seconds until power supply discharges as observed on the power 

supply LED indicator will turn off.

4.  Reconnect power supply to both AC to main voltage and DC connector to 

light source.

5.  Turn ON light source by rotating knob clockwise until clicks on and LED 

indicator light is on.

6.  Insert fiber cable completely to deactivate fiber sense interlock feature and 

ensure turret is rotated to position the fiber cable in front of either the UV or 

visible LED. You will feel a click when it is in the correct position.

7.  Rotate knob to increase light output intensity to desired output.

1.  Turn the LED Light Source off and unplug the power cord from both the wall 

outlet and the rear of the unit.

2.  Wipe the external surfaces clean with a cloth dampened with mild soap and 

water. 

DO NOT IMMERSE.

3.  Wipe the power cords clean with a cloth dampened with mild soap and water. 

DO NOT IMMERSE. DO NOT RECONNECT WET.

4.  DO NOT plug the power source into a wall outlet until it is thoroughly dry.
5.  It  is recommended to periodically clean the optical surface near the LED. 

Please use a soft cotton swab dipped in Isopropyl Alcohol and wipe the optic 

surfaces and allow to thoroughly dry prior to use.

For repair information, please contact Customer Service at:

Telephone:  (714) 236-8600

Email:  [email protected]

In the event the unit stops functioning, try the following steps to operate light 

source. The power supply exhibits internal protection circuitry for user safety  

precautions and will shutdown during certain instances. This equipment has been 

tested to ESD conditions according IEC 61326-1 and performs to performance 

criterion C. This means, that under certain conditions the overvoltage protection 

of the power supply may turn the power supply output and the unit off to prevent 

damage to the unit. In such case the power supply must be disconnected from 

main power to reset this fault condition. In the event the unit suddenly turns off, 

turn the unit off. Unplug the power supply from mains voltage (120V/ 230V). Wait 

~5 seconds and plug the power supply back into mains voltage. Turn the unit on.

Troubleshooting Suggestions

Repair

For repair information, please contact our team at:

Phone: Office: 908-496-8008

Email: [email protected]

Summary of Contents for iShot UV/White LED Light Source Kit

Page 1: ...UV White LED Light Source Kit...

Page 2: ...ironment 8 Electrical Ratings 8 Dimensions 9 Illumination System 9 Product Ordering Information 9 Approvals 10 Warranty 10 Agency Compliance Statements 11 FCC Class A Compliance Statement 11 Canadian...

Page 3: ...nsport Temperature Alternating Current Direct Current Manufacturer Do Not Immerse in Any Liquid Do not dispose of this product as unsorted municipal waste Prepare this product for reuse or separate co...

Page 4: ...e de lumi re DEL directe des fins m dicales WARNING MISE EN GARDE Unit MAY BE HOT Allow to cool before handling L appareil PEUT TRE CHAUD Il est important de le laisser refroidir avant d y toucher CAU...

Page 5: ...the proper position 5 Turn the power switch to the ON position LED indicator light will turn on when light source is powered Figure 3 6 Adjust the intensity control to set the light intensity to the...

Page 6: ...ruct air flow paths This device is designed to have proper forced air cooling paths to maintain thermal stable operation 2 Place in an area that provides adequate ventilation to prevent unit from over...

Page 7: ...and water DO NOT IMMERSE DO NOT RECONNECT WET 4 DO NOT plug the power source into a wall outlet until it is thoroughly dry 5 It is recommended to periodically clean the optical surface near the LED P...

Page 8: ...non condensing Storage Temperature 10 F to 140 F 20 C to 60 C Humidity 30 to 75 rh Atmospheric Pressure 700 hpa to 1060 hpa Mode of Operation Continuous Safety System Classification Class II System Po...

Page 9: ...hermal operation Dimensions Illumination Source Length 127MM 5 0 Width 90MM 3 5 Height 68MM 2 7 Weight 0 45 kg 1 0 lbs Product ID Description Order Code ULB 35rvi 35W RVI UV LED Light Source ACMI 1004...

Page 10: ...ifications when subject to normal use and service for a period of one year from the date of purchase from USHIO America or an authorized agent USHIO America will either repair or replace any component...

Page 11: ...al may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct t...

Page 12: ...Rev1 091517 MADE IN USA www intertest com 303 State Route 94 Columbia NJ 07832 USA Office 908 496 8008 Toll Free U S 800 535 3626 Fax 908 496 8004 Sold By...

Reviews: