CP MULTITHERM D.A HI

HOM161009163Ien-1 • 01/06/2017

Summary of Contents for CP MULTITHERM D.A HI

Page 1: ...Operating Instructions CP MULTITHERM D A HI HOM161009163Ien 1 01 06 2017...

Page 2: ...contents 1 General 3 2 Use 3 3 Explanation of symbols for warnings and signal words 3 4 Safety information 3 5 Installation accessories 4 6 Technical Data 4 7 Installation 4 8 Connection 5 9 Commissi...

Page 3: ...The SMART HEATER HI SYSTEM is used for demand ing heating tasks in explosive areas to maintain high temperatures in analytical instruments The temperature controller TC connected to the electrical hea...

Page 4: ...2 hexagon bolts 2 T head bolt type 28 15 M 6x15 4 M6 flat square nut 2 Washer shape A 6 4 4 Lock washers shape I 6 4 2 M6 hex nut If one of the components listed here is not included in the scope of d...

Page 5: ...es must be adhered to The connection cable must be permanently installed up to the inlet into the on site junction box taking the per missible bending radius 5 x outer diameter into consid eration In...

Page 6: ...versible damage to the heater and this would mean that the explosion protection can no longer be warranted The illustration may deviate from the heater 1 2 3 4 5 6 7 8 9 10 11 12 13 L N PE NTC RS 485...

Page 7: ...from bracket and remove it 13 Disposal Disassemble the components of the product taking the applicable local labour protection and environmental reg ulations into consideration and make sure that the...

Page 8: ...ir ectly contact INTERTEC Hess GmbH Raffineriestr 8 93333 Neustadt Donau Germany Phone 49 9445 9532 0 e mail info intertec info Website www intertec info Warranty The legally defined warranties and wa...