Spiromatic test

User Manual

99056C01

ENGLISH

Summary of Contents for Spiromatic test

Page 1: ...Spiromatic test User Manual 99056C01 ENGLISH...

Page 2: ...by copyright All rights are reserved Interspiro Oxydive and Divator are registered trademarks belonging to Interspiro This publication may not be copied photocopied reproduced translated or converted...

Page 3: ...Table of Contents Important information 5 Product description 6 Maintenance 10 Before use 11 After use 16 Depressurize the test equipment 16 Storage 17 Spiromatic test 3...

Page 4: ......

Page 5: ...l errors inaccuracies of current information or improvements and changes of equipment may be made at any time without prior notice Always refer to www interspiro com for product or document updates an...

Page 6: ...ssure regulator 4 4 Pressure regulator for a maximum permitted primary pressure of 300 bar 4351 psi Fitted with a safety valve that opens when the secondary pressure rises above 150 160 bar 2175 2320...

Page 7: ...sure of the pressure regulator 4 17 Test adapter housing to connect the test adapter in 18 Hose with cover for connection between the hose nipple and the test adapter 19 Adapter hose to connect betwee...

Page 8: ...ber and calibration date Technical data Material Aluminium anodized brass chromated plastic stainless steel Weight 12 kg Dimensions Length 475 mm Width 355 mm Height 175 mm Pressure maximum connecting...

Page 9: ...HP MP LP 1 HP High pressure area 2 MP Medium pressure area 3 LP Low pressure area Spiromatic test 9...

Page 10: ...ce and calibration every year If the Spiromatic test has not been calibrated by Interspiro for over a year it must not be used Strikes and or impacts can damage unprotected threads which can make it i...

Page 11: ...N 12021 Preparation Before doing any tests with the Spiromatic test it has to be prepared and it s fuctions tested 1 Make sure that the pressure gauge is on zero If needed turn the adjusting screw wit...

Page 12: ...mounted in the connecting piece 4 Turn the shutting off screw clockwise until it is closed This will stop the flow to the hose nipple 5 Make sure that the needle valve is closed If needed turn clockwi...

Page 13: ...essure of 100 bar 8 Slowly open the air supply 9 Turn the regulating screw in clockwise until the primary pressure gauge shows 80 bar NOTE If the pressure rises after the adjustment open the needle va...

Page 14: ...pply pressure gauge does not show any decrease in pressure during 60 seconds If the pressure drops the Spiromatic test is defect and must be repaired by Interspiro 11 Open the needle valve to vent the...

Page 15: ...13 Turn out the regulating screw until it feels loose counterclockwise Stop before it falls out 14 Follow the test instructions for testing each apparatus Spiromatic test 15...

Page 16: ...e the test equipment 1 Close the air supply 2 Turn out the regulating screw until it feels loose counterclockwise Stop before it falls out 3 Press in the selector valve until the system is depressuriz...

Page 17: ...orage Store in a cool dry and dust free environment Protect against direct sunlight UV radiation and direct heat 1 Make sure the system is depressurized 2 Assemble the protective plug Spiromatic test...