SECTION 4: PULSE OXIMETRY (OPTIONAL FEATURE)
Part No. 715-0080, Rev. B
- 44 -
PULSE OXIMETER SET-UP INSTRUCTIONS
The following steps will instruct the user how to initially set up the pulse oximeter and program low and
high alarm settings for SpO
2
and pulse rate. Additionally, when using the Masimo pulse oximeter, the
user can also set the algorithm mode, set the averaging mode, and view the oxygen perfusion index.
CAUTION:
Check alarm limits each time the system is used to ensure that they are appropriate
for the patient being monitored.
INITIAL SET-UP
1.
Turn the incubator power on using the Power Button (Figure 3 - 1, item 32).
2.
Connect the pulse oximeter cable to the connector on the side of the incubator
(Figure 3 - 2, item 2). Connect pulse oximeter sensor to the cable and to a suitable site
on the patient being monitored.
3.
To turn on the pulse oximeter displays, press the SET button (Figure 3 - 1, item 24) one
time and then either the up or down arrow button (Figure 3 - 1, item 25) until the %SPO2
screen displays “On”. If the screen shows “---”, there is no value available.
4.
When the pulse oximeter is disabled, the %SPO2 and PULSE screens will display “OFF”.
All alarms related to the pulse oximeter will also be disabled.
SETTING HIGH AND LOW %SpO
2
ALARMS
NOTE:
The pulse oximeter display will convert back to Normal Operation Mode after 5
seconds if no buttons are pressed.
1.
Press the SET button (Figure 3 - 1, item 24) three times (or until the %SPO2 HI LED
illuminates) to set the high SpO
2
alarm.
2.
Use the up or down arrow buttons (Figure 3 - 1, item 25) to set the high alarm to the
desired setting.
a.
Must be at least 1% higher than the SpO
2
low alarm setting.
b.
Can be set from 85 - 100%.
3.
Press the SET button four times (or until the %SPO2 LO LED illuminates) to set the low
SpO
2
alarm.
4.
Use the up or down arrow buttons to set the low alarm to the desired setting.
a.
Must be at least 1% lower than the SpO
2
high alarm setting.
b.
Can be set from 0 - 84%.
Summary of Contents for AirBORNE A750i
Page 1: ...A750i Transport Incubator Service Manual...
Page 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Page 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Page 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Page 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Page 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Page 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Page 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Page 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Page 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Page 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Page 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Page 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Page 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...