background image

3

Reed Switch Setting

Adjusting the Shock Sensor

DIP Switches

The following describes the DIP switch functions:

DIP Switches 1 and 2—adjust the Pulse Count.

DIP Switches 3 and 4—adjust the sensitivity setting of 
Gross Attack detection.

DIP Switch 5—enable/disable reed switch.

DIP Switch 6—not used.

Note 
In order for the LED to indicate shock detection while 
adjusting the sensitivity, be sure the reed switch is dis-
abled (DIP switch 5 OFF) or that the magnet is lined up 
with the reed switch if DIP switch 5 is ON.

Gross Attack Adjustment

1. To adjust the sensor for Gross Attack, set DIP switches 

1 and 2 to the ON position. This disables the Pulse 
Count so that the unit can only be activated by a Gross 
Attack.

2. Apply high level shocks to the mounting structure, using 

the LED as a guide to when the alarm trips (LED on for 
4 seconds). 

The LED will blink for 1 second every time the sensor 
detects a pulse. A shock that is severe enough to cause 
an alarm will cause the LED to light for approximately 4 
seconds.

3. Use switches 3 and 4 to adjust the Gross Attack sensitiv-

ity of the sensor (see Table 1).

4. Repeat step 2 each time you make a sensitivity change.

Pulse Count Adjustment

1. Set the sensor to the desired Pulse Count (see Table 2).

Note 
Pulse Count signals are counted at 1-second intervals 
and stored in a 30-second digital memory. These small 
signals can detect an intruder gently prying open a win-
dow or door frame.

2. To test the pulse count setting, generate small shocks on 

the mounting structure. Each time a shock is detected, a 
pulse is registered in memory and the LED will blink for 
one second. If the programmed pulse count is reached 
within the most recent 30 seconds, the alarm will trip 
and the LED will light for approximately 4 seconds. If 
the alarm trips for any reason, the stored pulses are can-
celled.

3. Use switches 1 and 2 to adjust the Pulse Count.

4. Repeat step 2 each time you make a sensitivity change.

Reed Switch Setting

After adjusting the sensor sensitivity, set DIP switch 5 to the 
appropriate setting, ON for enabled or OFF for disabled.

System Programming

This section describes the basic steps for adding the sensor 
to panel memory. Refer to the specific panel installation 
instructions for complete programming details.

The reed switch must be enabled and open when learning 
the sensor.

1. With the cover on the sensor, set the panel to Program 

mode.

2. Proceed to the Learn Sensors menu.

3. Select the appropriate sensor group and sensor number 

assignments.

4. When prompted by the panel to trip the sensor, remove 

the sensor cover to activate the tamper switch.

5. Exit program mode.

RF Testing

This section describes the basic steps for testing the sensor. 
Refer to the specific panel or receiver installation instruc-
tions for complete testing details.

1. Set the panel to Sensor Test.

2. Trip the sensor.

3. Listen for appropriate response from system sirens.

4. Exit Sensor Test.

Battery Replacement

When the system indicates that the sensor has a low battery, 
remove the old battery and install a new battery (Duracell 
DL123A or a Sanyo CR123A) into the battery holder, 
observing proper polarity.

Caution!
Replace only with a Duracell DL123A battery or a Sanyo 
CR123A battery. Observe polarity when installing a new 
battery. Installing the battery backwards may cause 
damage to the sensor.

Dispose of used batteries according to the manufacturer’s 
instructions and/or local government authorities.

 Table 1: Gross Attack Sensitivity Settings

DIP Switch 3 DIP Switch 4

Sensitivity

OFF

OFF

1 (maximum sensitivity)

ON

OFF

2

OFF

ON

3

ON

ON

4 (minimum sensitivity)

 Table 2: Pulse Count Adjustment

DIP Switch 1 DIP Switch 2

Pulse Count

OFF

OFF

4

ON

OFF

6

OFF

ON

8

ON

ON

Disabled

Summary of Contents for 60-886-95

Page 1: ...the intended location to make sure that the panel can receive sensor signal transmissions The sensor is an RF device and there may be blind or non operational locations within the installation Normally these can be overcome by moving the sensor or receiver Always mount the shock sensor so that the detector is on the frame and not on glass solid or hollow core doors See Figure 2 for mounting locati...

Page 2: ...slot at the top end of the unit and remove the cover see Figure 5 Figure 5 Remove the sensor cover 2 Using the flathead mounting screws secure the base to the mounting surface either vertically or horizontally as required see Figure 6 Figure 6 Sensor base mounting holes 3 Position the shock element and press it firmly into its socket see Figure 7 4 If using the reed switch use the two remaining sc...

Page 3: ...ely 4 seconds If the alarm trips for any reason the stored pulses are can celled 3 Use switches 1 and 2 to adjust the Pulse Count 4 Repeat step 2 each time you make a sensitivity change Reed Switch Setting After adjusting the sensor sensitivity set DIP switch 5 to the appropriate setting ON for enabled or OFF for disabled System Programming This section describes the basic steps for adding the sen...

Page 4: ...n air FCC Notices This device complies with FCC Rules Part 15 Operation is subject to the following two conditions 1 This device may not cause harmful interference 2 This device must accept any interference that may be received including interference that may cause undesired operation Changes or modifications not expressly approved by Interlogix Inc can void the user s authority to operate the equ...

Reviews: