E.
Trigger Send value
i.
Actual Values
ii.
Maximum Values
iii.
Minimum Values
iv.
Mean Values
F.
GP Output
– “General Purpose” Output
G.
Threshold Switch
4.3.3.
Threshold switch Mode
– Only Activated if Threshold Switch is selected in I/O type.
A.
Hysteresis switch (normal)
– Digital output becomes active if measuring value of
corresponding channel is above ON-threshold. It becomes inactive if measuring value of
corresponding channel is blow OFF-threshold.
B.
Window comparator
– Digital output becomes active if measuring value of corresponding
channel is between upper and lower threshold, otherwise inactive.
4.3.4.
Line Inverted
A.
Not inverted
– Active level is logical high = 5V. Inactive logical low is 0V.
B.
Inverted
– Active level is logical low – 0V. Inactive logical high is 5V.
4.3.5.
Default output level
– Level which digital I/O will output by default. That applies to all DIO
output types after power-on, before a set output condition occurs.
A.
E.g. set output level command if GP output type.