background image

 

 

8104738 Rev. 2 – 2019/03 

 

Page 

2

 of 

2

 

PLACE TRANSDUCER/S 

Place the probe or insert earphones in the infant’s ear/s or place the headset over the infant’s ears. 

 

 

 

The probe light will turn green when a good seal is obtained. 

 

RUN TEST 

Press the START button on the Titan or press the preamplifier button. 

 

PATIENT NOISE (EEG) 

After starting the test, the patient noise/EEG bar is displayed at the top of the screen.   

 

 

If the patient noise goes above the black line, try to calm the 

infant to reduce movement, crying, sucking etc. 

 

If the infant appears calm and the patient noise is not ideal, 

stop the test and select a protocol with a higher rejection 

level (if available).  

 

If the patient noise/EEG signal is still not ideal, refer to the 

EEG troubleshooting section below. 

RESULTS 

PASS result 

REFER result 

 

 

 

 

PASS – when the result reaches 100% for the pass criteria, PASS is displayed in green above the 

measurement. 

REFER – when the result does not reach 100% for the pass criteria within the measurement time, REFER is 

displayed in amber above the measurement.  

INCOMPLETE – if the test is stopped before a PASS or REFER is generated by the system, INCOMPLETE 

is displayed above the measurement indicating that the full test was not completed.   

 

EEG TROUBLESHOOTING 

When the patient noise/EEG signal is poor, and the patient is calm, try the following: 

  Turn off (if possible) all electronic devices/sources, such as lights, computers, mobile phones etc.  

  Ensure that the Titan battery is not being charged (via the cradle) during testing 

  If impedances were initially poor, clean the skin/use conductive gel and replace the electrodes 