background image

 

D-0128226-A – 2020/12 
Air Fx Caloric Irrigator – Instructions for Use - US 

 

Page 

10

 

 

Cool:

 

24°C

 

24. 1°C

 

 Irrigating

 

  

0:08  

NOTICE

 

The patient's perceived spinning sensation experienced during caloric irrigation is 

the desired result of stimulating a functional ear with a caloric irrigator. The 

clinician should remain at the patient's side to calm and reassure the patient that 

the spinning will pass in a minute or so. Nausea is an undesirable side effect 

which is temporary and not experienced by every patient. If the patient becomes 

nauseated, discontinue testing until the nausea has passed. Be prepared to 

provide the patient with an emesis basin. Stop all caloric testing for that visit if the 

patient vomits. 

 

 

 

3.7  Safety 

The Air Fx air irrigator uses 

distilled or demineralized water 

to maintain the temperature in the cooler / 

heater element. While the temperature of the coolant is not displayed, it is monitored. If the coolant 

temperature goes over a certain limit, the unit will stop operation and an “Over Temp Error” message is 

displayed. This error could be a result of insufficient water in the water reservoir or bubbles in the water 

line. Operating the irrigator in very cold (<10°C conditions can also cause this error). Make sure the 

irrigator has warmed up slowly to room temperature before operating.  

Over Temp Error

 

Call TechSupport 

 

The air delivery temperature is also monitored for safety. If this temperature exceeds 50°C, the irrigator 

will stop operation and an “Air Temp Error” message will be displayed.  

Air Temp Error

 

Call TechSupport 

 

If either of these conditions occurs, then shut down the irrigator and isolate the problem (ambient 

temperature too low, bubbles in the water line, 

distilled or demineralized water 

level too low etc. ). 

Contact Interacoustics’ Service Department or the local service representative if the problem cannot be 

resolved.  

 

Set the Air Fx irrigator near the edge of the table or cart in order for the hose to hang down, allowing the 

handle to rest properly in its cradle. When the irrigator is not in use, coil the Air Fx handle tubing using the 

hose management tape while still allowing the handle to be secured in the cradle.  

 

 

3.8  Turn off the irrigator 

The Air Fx will go into standby mode after ten minutes has elapsed or the irrigation temperature selection 

button is pressed twice. If the Air Fx is used in conjunction with the compatible VNG/ENG software, then 

the software will send the Air Fx into standby mode at the end of the test. When the Air Fx is in the 

standby mode, it is safe to turn off the power switch on the back panel.  

 

 

3.9  Draining the water reservoir 

The Air Fx should have the water reservoir on the top of the irrigator filled with 

distilled or demineralized 

water

. If the 

distilled or demineralized water 

needs to be drained prior to shipment or flushed before 

adding new 

distilled or demineralized water

, then the Air Fx should be drained according to the 

following instructions. Do not store irrigator in a location below 0°C (32°F).  

1.  Remove the water reservoir cap on the top of the irrigator.  

2.  Connect the supplied plastic syringe and tubing to the Fill / Drain Water Outlet on the back of the 

irrigator. Push the syringe plastic tubing fully into the Fill / Drain Water Outlet connector, and then 

gently pull on the tubing to verify the connection is secure and will not leak.  

3.  Switch the Fill / Drain Valve to the Fill / Drain position.  

Summary of Contents for Air Fx

Page 1: ...Instructions for Use US Air Fx Caloric Irrigator Science made smarter 8525533 D 0128226 A 2020 12...

Page 2: ...thout notice No part of this document may be reproduced or transmitted in any form or by any means without a prior written permission of Interacoustics A S FireWire is a registered trademark of Apple...

Page 3: ...re 9 3 6 Perform the irrigation 9 3 7 Safety 10 3 8 Turn off the irrigator 10 3 9 Draining the water reservoir 10 4 MAINTENANCE 12 4 1 General cleaning procedure 12 4 1 1 General Precautions 12 4 1 2...

Page 4: ...the caloric test with the correct irrigation temperature All personnel who operate the Air Fx should familiarize themselves with the contents of this manual prior to using the Air Fx with a patient A...

Page 5: ...C 10 Amperes Speculum 2 75 mm Speculum 4 25 mm 60cc syringe and tube USB Cable 3m 9 8ft User s Manual Drain Fill Kit Hose Management Tape 1 4 Warnings and precautions Throughout this manual the follow...

Page 6: ...he invoice serial number and a detailed report of the issue For any on site service related information please contact your local distributor If the system components are to be returned for service pl...

Page 7: ...sed at close distances to the equipment Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Performance Conditions...

Page 8: ...ons for Use Observe precautions for handling electrostatic sensitive devices It is against the law to dispose of electronic devices in the trash The crossed out wheelie bin symbol shows that the compo...

Page 9: ...ol Irrigation to be performed returns to standby state C Adjust temperature set point up 1 C D Adjust temperature set point down 1 C E Hose connection F Handle support G LCD output screen Figure 2 Rea...

Page 10: ...e tube to the Fill Drain Water Outlet Fully push the tube into the Fill Drain Water push in connector about 1 2cm of tubing Gently pull on the tubing to confirm the tube is secure 5 Switch the Fill Dr...

Page 11: ...urn on the irrigator Turn on the power switch on the back panel The Air Fx will initialize show the firmware version and then go into the standby state Select Cool Warm Air Fx v1 4 3 4 Attaching the s...

Page 12: ...mperature e g 23 4 C Cool 24 C 23 4 C Irrigation will be available after the desired temperature has been reached and stabilized Once this occurs the Air Fx will emit an indicator tone and will displa...

Page 13: ...n the irrigator and isolate the problem ambient temperature too low bubbles in the water line distilled or demineralized water level too low etc Contact Interacoustics Service Department or the local...

Page 14: ...ube Squirt any distilled or demineralized water into a waste container Reconnect the syringe to the tube Repeat this process until all of the distilled or demineralized water is removed from the inter...

Page 15: ...ion control The WHO s recommended level of decontamination for non critical devices is cleaning Disinfection and sterilization are not recommended However in case of an epidemic outbreak disinfection...

Page 16: ...r Then fill the irrigator with distilled or demineralized water and run couple of cool irrigation cycles Drain the irrigator to ensure that the disinfectant is washed out Then refill the reservoir wit...

Page 17: ...ed Interacoustics service representative altered in any way so as in Interacoustics judgement to affect its stability or reliability subject to misuse or negligence or accident or which has had the se...

Page 18: ...ctric and electronic waste as unsorted municipal waste Electric and electronic waste may contain hazardous substances and therefore has to be collected separately Such products will be marked with the...

Page 19: ...D 0128226 A 2020 12 Air Fx Caloric Irrigator Instructions for Use US Page 16...

Page 20: ...faster 3 m 9 8 ft cable Case dimensions 35W x 32D x 22H cm 13 8 x 12 6 x 8 7 in Air hose length 3 m 9 8 ft Weight 8 kg 18 lb Voltage 110 130 VAC 50 60 Hz 220 240 VAC 50 60 Hz Fuses 2 each 110VAC unit...

Page 21: ...ons or decreased electromagnetic immunity of this equipment and result in improper operation The list of accessories and cables can be found in this section Portable RF communications equipment includ...

Page 22: ...omagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Air Fx can help prevent electromagnetic interferences by maintaining a minimum distance between p...

Page 23: ...nvironment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or residential environment Vol...

Page 24: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Page 25: ...s or fax No to whom Interacoustics may confirm reception of the returned goods The above mentioned item is reported to be dangerous to patient or user 1 In order to ensure instant and effective treatm...

Reviews: