GUIDANCE AND MANUFACTURER’S DECLARATION — ELECTROMAGNETIC IMMUNITY
InTENSity at Home
™
electrical stimulators are intended for use in the electromagnetic environment specified below.
The customer or the user of these electrical stimulators should assure that it is used in such environment.
Immunity test
IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF
lEC 61000-4-6
Not applicable
Portable and mobile RF Communications equipment should
be used no closer to any part of the device, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the Transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated RF
lEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m