3.2 CALIBRATION
Dealer Calibration Procedures:
1. Turn scale off.
2. Press and hold TARE, then turn scale on.
3. Scale displays F1
4. Press MODE
5. Scale displays offset value
6. Press and hold MODE, Scale displays YES.
7. Scale will self calibrate zero point before proceed to the first
point calibration.
8. After zero point calibration is done, scale displays LOAD XXXX or
XXXX
9. Load calibration load as request.
10.
Wait until scale displays LOAD XXXX(second point).
11.
Load calibration load as request or press MODE to abandon second
point calibration, wait until scale displays DONE.
12.
Calibration completed and scale is ready for operation.
Auto Calibration Procedures:
1. Turn scale off.
2. Press and hold MODE, then turn scale on.
3. Scale displays CAL.?
4. Press MODE again, Scale displays YES.
5. Scale will self calibrate zero point before proceed to the first
point calibration.
6. After zero point calibration is done, scale displays LOAD XXXX or
XXXX
7. Load calibration load as request.
8. Wait until scale displays LOAD XXXX(second point).
9. Load calibration load as request or press MODE to abandon second
point calibration, wait until scale displays DONE.
10.
Calibration completed and scale is ready for operation.
Note
:
Auto calibration can only be performed when the span
value Is /-10% tolerance of dealer calibration.
3.3 DISABLE CALIBRATION AND INTERNAL FUNCTIONS SETTING WITH JUMPER
SW.(JP1)
The JP1 is used to control calibration and internal functions
setting(F1 and F3). move this jumper to CAL-LOCK position to disable
calibration and internal functions setting.
Summary of Contents for AXM-1500
Page 5: ...5 2 3 GENERAL SPECIFICATION 2 3 1 Overall View...
Page 15: ...15 5 ELECTRICAL CIRCUITRY 5 1 SCHEMATICS...
Page 16: ...16...
Page 17: ...17...
Page 18: ...18...
Page 19: ...19...
Page 22: ...22 AXM 40 1 TOP LAYER AXM 40 1 BOTTOM LAYER AXM 40 1 TOP OVERLAY AXM 40 1 BOTTOM OVERLAY...
Page 27: ...27 7 APPENDIX...