IC8P Battery Charger/Analyser
Operator Manual
7.3
Entering Discharge Data
From the
MAIN
screen by selecting
CHARGE
->
MANUAL
From the
MAIN
screen by selecting
CHARGE
->
LIBRARY
->
->
MODIFY
From the
MAIN
screen by selecting
PROCESS
->
PROCESS #
->
VIEW/EDIT
->
MODIFY
->
CHARGE BATTERY
.
The screen will request
CHARGE TYPE
.
Select
>> NEXT >>
or
<< PREV <<
to change to either : -
CAPACITY TEST TO TIME
CAPACITY TEST TO TARGET
FULL DISCHARGE
The following tables explains what the values entered mean.
CAPACITY TEST TO TIME
Value Required
Units
Description
Nominal Volts
Volts.
The battery nominal voltage is used to determine what
power rating is required for discharge and also the
help validate the data entered.
Target Volts
Volts.
The target volts is the discharged voltage that the
charge should stop the discharge and report the
battery capacity.
Discharge Amps
Amps.
This is the constant current discharge amps that the
battery has to be discharged at.
Discharge Time
Hours and
Minutes.
This is the discharge time (usually 1 hour). When this
time is reached the charge will stop and report the
battery capacity of 100%.
CAPACITY TEST TO TARGET
Value Required
Units
Description
Nominal Volts
Volts.
The battery nominal voltage is used to determine what
power rating is required for discharge and also the
help validate the data entered.
Target Volts
Volts.
The target volts is the discharged voltage that the
charge should stop the discharge and report the
Doc: DWG1062-06-R3 IC8P Operators manual.odt
Page 28 of 46