TABLE VI.
Troubleshooting Guide
MESSAGE CODE
Description/Symptom
PROBA BLE CAU SE
ACTION
M SG-0
Comm unications not
being received from
probe(s)
1.
Improper cable hookup
2.
Blown m ain fuse
3.
Failed RS -485 circuit
4.
Damaged flow sensor
1.
Verify plug-in connector properly mated
2.
Check all cable connections
3.
Che ck F1 fus e on prob e board
4.
Con tact factory
M SG-1
Invalid data received
from probe(s)
1.
Failed RS-485 communications
component
1.
Check wiring
2.
Con tact factory
M SG-2
Flow sensor heater
“OFF”
1.
Blown h eater fuse
2.
Failed electronic component
1.
Con tact factory
M SG-3
Relative saturation
sensor power “OFF”'
1.
Probe temperature too high
1.
Ch eck to be su re prob e tem peratu re is
<160 °F
2.
Con tact factory
M SG-4
RS heater “FF”
1.
Com ponen t failure
1.
Con tact factory
M SG-5
Pressure sensor heater
“FF”
1.
Problem with pressu re port
circuitry
1.
Con tact factory
M SG-6
Tempera ture alarm
(above 210°F/99°C)
1.
Steam in exhaust pipe
1.
Remove probe or cool line
ASAP!
On ce line h as co oled d ow n and probe is
reinstalled, check unit for proper function
M SG-7
Wet probe
1.
This usually indicates a condenser
design or operation problem that
requires corrective action
2.
Hogging mode during start-up
1.
Con tact factory
M SG-8
RS sensor problem
1.
RS sensor has failed
2.
Con tact factory
Mass flow output
saturates
high
, will not
respond to flow changes
1.
Flow rate is not within range of
calibration
2.
Blown h eater fuse
3.
Failed electronic component
1.
Contact factory about re-ranging
instrument
2.
Replace fuse
3.
Con tact factory
Mass flow output
saturates
low
, will not
respond to flow changes
1.
Flow rate is not within range of
calibration
2.
Failed electronic component
1.
Contact factory about re-ranging
instrument
2.
Con tact factory
When using the
RheoVac
software on a PC, the software may give you a “Communication Error or
Message.” Use the Table VII to determine the source of this message and appropriate action.
Note: If the
RheoVac
instrument is operating without message indications but output readings are
questionable, please send to Intek by telefax or e-mail the following plant data: turbine back pressure,
hot well temperature, load, and inlet and outlet circulation water temperatures, along with at least 24
hours of data from the
RheoVac
instrument/probe for the same time period.